Idiomas disponibles

  • ESPAÑOL, página 39
5. Supply the specified values (Table 19). Make sure that
the error is in the specified limits.
Table 19: RTD (Ohms) Output Error Limits
Excitation
(mA)
Ohms (Ω)
0
0.50 to 3.0
100
0.50 to 3.0
200
0.50 to 3.0
300
0.50 to 3.0
400
0.50 to 3.0
1000
0.05 to 0.8
2000
0.05 to 0.4
4000
0.05 to 0.3
a. Refer to "Specification".
4.10 Procedures: TC (mV) Input/Output
1. Connect the instrument to the calibration equipment:
TC (mV) input = Figure 9b
TC (mV) output = Figure 9d
2. Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3. Use the calibration menu (Table 9) to do the
calibration:
TC (mV) input = three-point calibration (-FS, Zero
and +FS).
TC (mV) output = two-point calibration (Zero and
+FS).
The display shows the applicable instructions to
complete the calibration.
4. To make sure that the calibration is correct, select the
applicable TC (mV) input or output task (Table 2) and
apply the necessary values:
TC (mV) input: -10, 0 (short circuit)
Then TC (mV): 25, 50, 75
TC (mV) output: -10, 0, 25, 50, 75
5. Make sure that the error is in the specified limits
(Table 20).
Table 20: TC (mV) Input/Output Error Limits
Calibrator Error TC
Input or
(mV)
Output TC
(mV)
-10
0.000 5
0.000 18
0
-
0.000 05
25
0.000 6
0.000 36
50
0.000 8
0.000 68
75
0.001 0
0.000 99
4.11 Procedures: IDOS UMM
Refer to the user manual for the IDOS UMM.
When the calibration is complete, the instrument
automatically sets a new calibration date in the UMM.
5. Specification
All accuracy statements include one year stability.
Copyright 2005 Baker Hughes Company.
14 | DPI880 Instruction Manual–English
Permitted
Calibrator
a
DPI880 Error
Error (Ω)
(Ω)
0.003
0.05
0.004
0.06
0.005
0.06
0.007
0.07
0.008
0.07
0.015
0.30
0.026
0.40
0.049
0.80
Permitted DPI880 Error
TC (mV)
0.008
0.008
0.006
0.006
0.010
0.010
0.014
0.014
0.018
0.018
5.1 General
Parameter
Value
Languages
English [Default]
Operating
14 to 122°F (-10 to 50°C)
temperature
Storage
-4 to 158°F (-20 to 70°C)
temperature
Humidity
0 to 90% without condensation (Def Stan
66-31, 8.6 cat III)
Shock/ Vibration
EN 61010:2010; Def Stan 66-31, 8.4 cat III
EMC
EN 61326-1:2013
Safety
Electrical - EN 61010:2010; CE & UKCA
Marked
Size (L: W: H)
7.1 × 3.3 × 2.0 in (180 × 85 × 50 mm)
Weight
15 oz (425 g)
Power supply
3 × AA alkaline batteries
Duration
Measure functions (area ): ≈ 60 hours Dual
Function, mA measure (area ): ≈ 7 hours
(24 V Source at 12 mA)
5.2 Electrical (Figure A1 - Item 10)
Parameter
Value
Range (Measure):
0 to ±55 mA
0 to 4000Ω
0 to ±120 mV
0 to ±30 V
Accuracy: Measure mA
0.02% of reading + 3 counts
Accuracy: Measure mV
0.02% of reading + 2 counts
Accuracy: Measure V
0.03% of reading + 2 counts
Range (Supply):
0 to 24 mA
0 to 4000Ω
0 to 120 mV
0 to 12 V
Accuracy (Supply): mA,
0.02% of reading + 2 counts
mV, V
Temperature coefficient
(Measure or supply)
14 to 50°F, 86 to 122°F
0.0017% FS / °F
(-10 to 10°C, 30 to 50°C)
(0.003% FS / °C)
Connectors (Figure A1 -
Four 0.16 in (4 mm) sockets
Item 10)
One TC mini-connector socket
a. Refer to "Resistance Ranges (Ohms/RTD)".