After two successive inspiration phases the
alarm should be set off. If audio re-
connection
sponse is enabled, the ventilator announces
"Check ventilation system and settings".
Drop in O
pressure (<2.7 bar O
2
1. Open the valve of the oxygen cylinder
2. Switch on MEDUMAT Easy.
3. Close the valve on the oxygen cylinder. When
the oxygen pressure in the system has fallen be-
low 2.7 bar, the
off.
If audio response is enabled, the ventilator an-
nounces "Check pressure hose system and gas
supply".
Power supply (
The alarm that indicates a failing battery is checked
automatically in the self test that runs when
MEDUMAT Easy is switched on.
The power supply is in order if no alarm is set off
when the valve on the oxygen cylinder is opened and
MEDUMAT Easy is switched on and starts to
function.
alarm should be set
<2.7 bar O
2
)
+ –
Functional checks
dis-
)
2
.
slowly
53
EN