3.2 Owner/operator and user qualification
3.3 Ventilation function
10
11
12
9
0
10
12
10
13
7
11
14
16
5
30
3
Freq.(min
)
-1
MV (l/min)
•
As an owner/operator or user, you must be familiar
with the operation of this medical device. Observe
the legal requirements for operation and use (in Ger-
many, the regulations governing owner/operators of
medical devices apply in particular). Basic recom-
mendation: get a person authorized by WEINMANN
Emergency to provide you with proper instruction
about the handling, use and operation of this medi-
cal device.
MEDUMAT Easy operates within a pressure range of
2.7 to 6 bar and at a flow rate of not less than
14
70 l/min O
15
It uses high-pressure, medicinal-grade oxygen. An
16
external pressure reducer brings this down to the re-
Demand-
flow
quired operating pressure. The oxygen supply is fed
in at input valve.
The ventilation settings are continuously variable.
These settings (frequency and volume per minute
are coupled) and the inspiration/expiration ratio of
1:1.67 are regulated by internal electronic control
mechanisms.
The gas for inspiration flows along the hose and
through the patient valve and either the mask or
tube into the patient's airways. The patient valve is
permits respiration-controlled oxygen inhalation
in Demand mode.
. It has a built-in power pack.
2
EN
Description
11