6.10 Audio response for user guidance
32
EN
Using the ventilator
Stenosis:
Disconnection:
< 2.7 bar:
:
+ –
Even if audio response is enabled, no messages will
be output for the periods stated. The visual alarm
will continue to flash.
If the cause of the alarm is not eliminated, the acous-
tic alarm will start to sound again after a short inter-
val. Audio response will also resume automatically.
Both the visual and acoustic alarms are cancelled au-
tomatically as soon as the malfunction is eliminated.
Selecting language / Switching off audio
guidance
The language setting can only be selected if the unit
is switched off.
To select a language or to switch off the audio re-
sponse facility, proceed as follows:
1. Hold down the mask/tube switch. Switch on the
ventilator at the ON/OFF switch.
2. Then release the mask/tube switch. The unit is
now in the language selection menu. The venti-
lation pressure gauge displays the most recent
language setting. The following languages are
assigned to the individual LEDs:
30 seconds
30 seconds
30 seconds
120 seconds