7. Hygienic preparation

7.1 MEDUMAT Easy

7.2 Patient valve

40
EN
Hygienic preparation
the MEDUMAT Easy and any acces-
After every use
sories used must undergo hygienic preparation.
Be sure to carry out a functional check after ev-
ery hygienic preparation (see "8. Functional
checks" on page 45).
This product may contain disposable items.
posable items are intended to be used only once.
So use these items only once and do
cess them. Reprocessing disposable items may
impair the functionality and safety of the prod-
uct and lead to unforeseeable reactions as a re-
sult of ageing, embrittlement, wear, thermal
load, the effects of chemical processes, etc.
You can keep MEDUMAT Easy clean by simply wip-
ing with disinfectant as described in section 7.6 .
Never immerse MEDUMAT Easy in disinfectants or
other liquids. Otherwise damage may be caused to
the unit, thus endangering users and patients.
1. Disconnect the patient valve from the hoses.
Always grasp the hoses by their ends. Otherwise you
might damage or tear them.
Dis-
repro-
not