Technical Data - Weinmann MEDUMAT Easy Descripción Del Aparato E Instrucciones

Ocultar thumbs Ver también para MEDUMAT Easy:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

12. Technical data

Dimensions
LxHxB in mm
Weight incl.
accessories
Device category
(93/42/EEC)
Operation:
Temperature range
Humidity
Air pressure
Storage
Electromagnetic
compatibility (EMC)
according to
EN 60601-1-2
and EN 794-3:
– Radio interference
suppression
– Radio interference
resistance
Control
Gas input
Operating pressure
Minimum gas
volume required
Insp-exp. ratio (I:E)
Ventilation frequency
Minute volume (MV)
MEDUMAT Easy
100x145x90
incl. connections
approx. 0.6 kg
II b
–18 °C to +60 °C
max. 95 %
(1)
70 kPa
to 110 kPa
–40 °C to +70 °C
Test parameters and limit
values can be requested
from the manufacturer
(WEINMANN Emergency
Medical Technology
GmbH + Co. KG,
Frohboesestraße 12,
22525 Hamburg,
Germany).
EN 55011 B
EN 61000-4 Parts 2 to 6,
Part 11
timing pulse, volume
constant
Medical oxygen
(2)
2.7 to 6.0 bar
70 l/min O
2
1:1.67
continuously variable
-1
from 10 to 30 min
continuously variable
from 3 to 16 l/min
MV tolerances:
Room temp. (20 °C)
–18 °C to +60 °C
Max. ventilation
pressure
O
concentration
2
Pressurized gas
connection
Ventilation hose
connection
Patient valve
– inspiration tube
– mask/endotracheal
tube
Patient valve
– expiration tube
Power supply
Life expectancy
Max. storage period
Auxiliary power for
alarm system
Max. storage period
Ventilation hose
Degree of protection
against water
Standards applied
(1) Under normal atmospheric conditions, 70 kPa
correspond to a maximum altitude for use of
approx. 3,000 meters.
(2) 1 bar = ˆ 100 kPa
(3) 1 mbar = ˆ 1 hPa
Technical data
MEDUMAT Easy
for 3 l/min = ±20 %
for >3 l/min = ±15 %
for 3 l/min = ±35%
for >3 l/min = ±20%
(3)
20 or 45 mbar
100 % O
(at least 98 %)
2
External thread G 3/8
External diameter 13 mm
15 mm socket
22 mm plug
ISO 5356-1
30 mm socket
ISO 5356-1
maintenance-free lithium
battery 3.6 V; 5.2 Ah,
> 2 years
10 years after delivery
Button cell CR2430
10 years after delivery
– Spiral silicone DN 10
– PVC hose DN 8,
disposable
IP54
EN 794–3; EN 1789
EN
63