ALARMS, SAFETY, AND SELF-DIAGNOSIS
Tables 4–6 show the behaviour of the alarm (buzzer) and ring back function when pressing the BUZZER
STOP key.
Table 4. In cases other than those covered in Tables 5 and 6
Remote Alarm setting
ON: Remote alarm setting not linked with
BUZZER STOP key
OFF: Remote alarm setting linked with
BUZZER STOP key
Note: Resolve the cause of the alarm, as the alarm itself is not deactivated by pressing the BUZZER
STOP key.
Table 5. In the case of high limit temperature alarm or Err38
Remote Alarm setting
ON: Remote alarm setting not linked with
BUZZER STOP key
OFF: Remote alarm setting linked with
BUZZER STOP key
Note: Close the outer door when Err38 is activated.
Table 6. In cases of Err01, Err11, Err12, Err18, or door alarm
Remote Alarm setting
ON: Remote alarm setting not linked with
BUZZER STOP key
OFF: Remote alarm setting linked with
BUZZER STOP key
*1: The remote alarm does not work for the door alarm.
Note: When Err01 is activated, connect the new CO
stop the buzzer. In addition, when the optional MCO-50GC is installed and the gas supply is switched to
the reserve gas cylinder, press the BUZZER STOP key and replace the gas cylinder.
ROUTINE MAINTENANCE
To use this unit in a clean condition, clean the chamber and all the inner attachments at least once a month.
1. Remove all the inner attachments by the procedures shown on page 22.
2. Clean the chamber and all the inner attachments by the procedures shown on page 21.
3. Install all the inner attachments by the procedures shown on page 23.
●When there is excessive dirt, contact our sales representative or agent.
82
Buzzer from CO
Ring Back
When pressing
setting
the BUZZER
STOP key
ON
OFF
OFF
(Alarm is not
ON
cancelled)
OFF
Buzzer from CO
Ring Back
When pressing
setting
the BUZZER
STOP key
ON
OFF
ON
ON
OFF
Buzzer from CO
Ring Back
When pressing
setting
the BUZZER
STOP key
ON
OFF
OFF
(Alarm is
ON
cancelled)
OFF
gas cylinder and press the BUZZER STOP key to
2
incubator
2
When pressing
When the Ring Back
the BUZZER
set time passes
STOP key
ON
OFF
ON
OFF (Alarm is
OFF
not cancelled)
incubator
2
When pressing
When the Ring Back
the BUZZER
set time passes
STOP key
ON
(Under continuation)
(Continue)
*1
incubator
2
When pressing
When the Ring Back
the BUZZER
set time passes
STOP key
OFF
(Alarm is already
(Alarm is
cancelled)
cancelled
Remote Alarm
When the Ring Back
set time passes
ON
ON
(Under continuation)
ON
OFF
Remote Alarm
When the Ring Back
set time passes
ON
ON
(Under continuation)
Remote Alarm
When the Ring Back
set time passes
OFF
OFF
(Alarm is already
*1
*1
)
cancelled
)