ELECTRIC LOCK (OPTION)
6. Move the cursor to select each digit of the password (6 digits) with the
Confirm Password
left/right keys (◁▷), set the numerical values using the up/down keys
(△▽), and then press the Enter key (ENTER).
0
*****
▶ The Confirm Password screen is displayed [Fig. 6].
7. For confirming the password, move the cursor to select each digit of
the password (6 digits) with the left/right keys (◁▷) again, set the same
<Right side of the display>
[Fig. 6]
numerical values using the up/down keys (△▽), and then press the
Enter key (ENTER).
▶ The setting is specified and the Door Lock screen is displayed.
*If the password in Confirm Password does not match the password that was first entered, the following
message is displayed on the right side of the display. Press either the up/down/left/right key to return to the
Door Lock screen, and then repeat the procedure from Step 4 above.
Confirm Password
Incorrect PW
Press Any Key
<Right side of the display>
Reference:
- This outer door unlock password is used when opening the outer door when it is locked. This
password differs from the key lock password (pages 46 – 47).
8. Press the menu key (MENU/HOME) to display the home screen.
The display returns to the home screen automatically when 90 seconds has passed without any key
operation.
Note: To prevent unauthorized use of the password, the password should be properly managed by only
qualified administrators.