INSTALLATION
Installation site
For correct operation of the incubator, install it in a location with the following conditions (referring also to
the precautions for safe operation on pages 5–9):
Install in an environment with normal air.
Do not expose to direct sunlight.
Do not install it near significant heat sources, such as heaters, boilers, ovens, or autoclaves.
Ensure ambient temperature is at least 5 C lower than set temperature.
Install in a site with a relative humidity of 80 %R.H. or lower.
Do not expose to flammable or corrosive gas.
Do not install where objects may fall on the incubator from above.
A flat surface where the floor is also capable of bearing the total combined weight (product +
optional accessories + stored items).
Do not install the incubator near vibration products.
16