Phcbi MCO-50AIC Instrucciones De Operación página 263

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

ALARMAS, SEGURIDAD Y AUTODIAGNÓSTICO
Código de error
Err 16: CO
Box Heater Error
2
Err 17: Heater SSR Open
Err 18: UV Lamp Abnormal
Err 20: Door Lock Failure
Err 39: Fan Motor Error
Err 56: Communication Failure
Err 65: Peltier Fan Error
Cuando está instalado un sistema UV opcional MCO-170UVS.
※1
Cuando está instalado un bloqueo eléctrico opcional MCO-170EL.
※2
Display OLED
Mensaje
Overheat alarm is activated,
or the sensor box heater
burnout occurs or the sensor
box heater SSR is
short-circuited.
Overheat alarm is activated,
or any heater SSR burnout
occurs.
The UV lamp is burnt out.
1
The outer door is open during
2
autolocked by electric lock.
Rotational speed of fan motor
is abnormal.
Communication between
display and control is
unstable.
Rotational speed of peltier fan
motor is abnormal.
Tipo de alarma
Zumbador
Error del calentador
de la caja del sensor
Desconexión de SSR
para calentador
Fallo de la lámpara
intermitente
UV
Error de bloqueo de
cerradura
Fallo del motor del
ventilador
Error de
comunicación
Error del motor del
intermitente
ventilador de Peltier
Operación
Alarma
de
remota
seguridad
-----
Tono
ON
Lámpara
UV OFF
La válvula
de CO
está
cerrada.
-----
-----
-----
Tono
ON
2
79
Tabla de contenido

Solución de problemas

loading

Tabla de contenido