Phcbi MCO-50AIC Instrucciones De Operación página 68

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

GAS AUTO CHANGER (OPTION)
When the CO
density level in the chamber remains unchanged even though the CO
2
is open, the unit assumes the currently connected CO
switches the CO
gas supply line to the other one. The behavior in that case is shown in Table 2.
2
1. When there is gas remaining in cylinder A, the unit operates using CO
cylinder A (Situation 1).
2. When the CO
2
gas is not supplied into the unit even though the gas valve in the unit is open (Situation 2).
3. When Situation 2 continues for 2 to 3 minutes, the gas supply line is automatically switched assuming
the gas cylinder to be empty. At this time, the CO
gas supply indicator A is displayed in reverse video and blinks (Situation 3).
4. Press the BUZZER STOP key to stop the CO
blinking (Situation 4).
5. Replace the empty gas cylinder A with a new one immediately after the Situation 4 (Situation 5).
6. When the cylinder B becomes empty, the supply line is again switched to the cylinder A.
Table 2. CO
gas supply line automatic switching
2
Situation
CO
gas is being
2
1
supplied from cylinder A.
CO
density in the
2
chamber does not
2
increase even if CO
2
gas valve is open.
CO
gas supply line is
2
3
automatically switched
to cylinder B.
The BUZZER STOP
4
key is pressed.
Empty gas cylinder A is
5
replaced with a new
one.
Notes:
When the BUZZER STOP key is not pressed in the Situation 4 and the CO
without the CO
gas cylinder A being replaced in the Situation 5, the switch operation between CO
2
supply lines A and B will be repeated. In this case, replace the both CO
ones immediately, and press the BUZZER STOP key.
The switching of CO
switching of CO
gas cylinders may also occur in the situations where the gas tube is clogged, the gas is
2
leaking, the gas pressure has dropped down, or the valve on the gas cylinder is insufficiently open etc.,
even though the CO
68
gas cylinder A is empty, the level of CO
CO
gas
2
Supply
Cylinder
line
A
A
Remaining
A
Empty
B
Empty
B
Empty
B
Remaining
cylinders is determined by an increase of CO
2
gas cylinder being used is not empty.
2
gas cylinder to be empty and automatically
2
density in the unit does not increase because
2
gas empty alarm is activated, the buzzer sounds, and
2
gas empty alarm. The gas supply indicator A stops
2
CO
gas supply line indicator
2
Cylinder
A
B
Reverse
video and
Remaining
light on
Reverse
video and
Remaining
light on
Blink
Remaining
Light on
Remaining
Light on
Remaining
gas valve in the unit
2
gas supplied from the CO
2
Message
B
display area
Light on
Light on
Reverse
Err01:
video and
CO
(and buzzer sound)
light on
Reverse
video and
light on
Reverse
video and
light on
gas cylinder B gets empty
2
gas cylinders A and B with new
2
density in the chamber. The
2
gas
2
-----
-----
Gas Empty
2
-----
-----
gas
2