BASIC PARAMETERS

Setting temperature

The chamber temperature during the normal operation of the incubator can be set as follows. The incubator
automatically starts operation using these settings after power-on.
 Settable range: 0.0
 Factory setting: 37.0
1. On the home screen, press the menu key (MENU/HOME).
▶ The left side of the display changes to the Menu screen.
2. When the cursor on the menu screen is on Set Value [Fig. 1], press
the enter key (ENTER).
▶ The display changes to the Set Value screen.
3. When the cursor on the Set Value screen is on Temperature [Fig. 2],
press the enter key (ENTER).
▶ The right side of the display changes to the Temperature screen, and
the current set value (37.0) is displayed [Fig. 3].
4. Move the cursor using the left/right keys (◁▷) and select the digit you
want to change, and then use the up/down keys (△▽) to change the
number.
5. Press the enter key (ENTER).
▶ The value is confirmed and the display returns to the Set Value
screen.
*If the entered setting value is out of the settable range, the following
message is displayed on the right side of the display. Press any key on
the control panel to change to the Temperature screen [Fig. 3].
6. Press the menu key (MENU/HOME) to display the home screen.
The display returns to the home screen automatically when 90 seconds has passed without any key
operation.
32
o
o
C – 50.0
C
o
C
Temperature
Invalid Value
Press Any Key
0.0°C – 50.0°C
<Right side of the display>
Menu
Set Value
Data Log
Lock
Alarms & Controls
<Left side of the display>
Set Value
Temperature
: 37.0°C
CO
ON/OFF
: ON
2
CO
Setting
: 5.0%
2
Overtemp
: 53.0°C
<Left side of the display>
Temperature
37.0
°C
0.0°C – 50.0°C
<Right side of the display>
[Fig. 1]
[Fig. 2]
[Fig. 3]