BEFORE STARTING OPERATION

Filling the humidifying pan

Use the following procedure to fill the humidifying pan with water or to replace water in it.
1. Pull out the humidifying pan (Fig. 1).
2. Dispose of the remaining water and clean the
humidifying pan with a diluted detergent. Then rinse
it thoroughly with distilled water and wipe it using
alcohol for disinfection.
3. Wipe all moisture from the bottom of the chamber.
4. Return the humidifying pan to the chamber and
pour sterile distilled water (approx. 1.5 L, preheated
to 37 C) (Fig. 2).
5. Place the humidifying pan flush against the rear wall in the chamber, and close the inner and outer doors.
Notes:
• Operation without water in the humidifying pan may temporarily increase the chamber temperature,
resulting in higher temperature than the set temperature.
• Preheat the water to 37 C. Adding cool water will lower the temperature and humidity in the chamber.
• Install the humidifying pan in a longitudinal direction with its shorter side placed at the back.
• Refill the humidifying pan with water promptly when the volume of water decreases.
• Mixing any reagent into the water for humidifying may have an adverse effect on the cultivation. Do not
add any reagent to the water particularly when using the UV lamp since the UV light may cause the
mixture to deteriorate.
• After cleaning, change the water in the humidifying pan. The alcohol may influence the correct control of
the operation.
• The humidity control bar in the duct is always kept in low temperature and the moisture in the chamber is
recondensed. Slide the humidifying pan down right under the humidity control bar, otherwise the
recondensed water droplets will fall directly to the bottom of the chamber and form a pool.
• When the pooled water evaporates, it may leave a white mark on the bottom. This is not abnormal. Wipe
it off with a piece of gauze or unwoven cloth moistened with alcohol for disinfection. If the mark cannot be
removed, scrub it off using a cleansing cream.
24
Fig. 1
Fig. 2
Tabla de contenido