CAUTION
NEVER place the hand piece in the charging unit without the battery inserted in the hand piece.
Insert the battery into the hand piece in the sequence described below:
1 Remove the battery from its packaging.
WARNING
Never short circuit the terminals of the battery with any metal object as it could cause
a risk of burns, fi re and explosion.
2 Use a flat tool (for example a coin) to unscrew counter-clockwise the battery cap of the
hand piece ( Fig . 1 and 2 ).
3 An arrow is marked on the battery case ( Fig . 3 ).
NOTE
This arrow indicates the orientation with which the battery must be inserted into the hand piece.
4 Slowly slide the battery with the arrow oriented toward the housing into the hand piece until it stops ( Fig . 4 ).
NOTE
GB
If the battery is inserted in the wrong way (i.e. in the opposite direction of the arrow),
the hand piece will not function. In case the battery is inserted in the wrong way,
no damage is caused to the hand piece's electronics.
- 12 -
WARNING
Use original Kulzer batteries only. The use of other manufacturer batteries or non-rechargeable batteries /
primary batteries is a potential hazard and can cause irreversible damage to the unit.
5 Use a flat tool (for example a coin) to screw clockwise the battery cap of the hand piece.
NOTE
The unit will not function properly if the cap of the hand piece is not screwed on completely.
5.4
Inserting the fi bre optic into the hand piece
• Rotate the light guide slightly ( 1 ) and insert it into the hand piece ( 2 ) until final positioning
is reached. The final position is reached when you hear a " click ".
• Mount the protective light cone ( 3 ) on the top end of the light guide.
Positioning the light guide
• Rotate the light guide into the desired position for the polymerisation.
• To make full use of the light intensity provided, place the light guide as close to the composite as possible.
Avoid direct contact with the composite material!
CAUTION – Keep the light guide clean at all times to obtain full light intensity.
The light intensity is considerably reduced if:
– The light guide is not in perfect condition (scratched or chipped).
– The light guide is not correctly placed in the hand piece.
– The light guide has composite stains.
WARNING
A damaged light guide reduces the light intensity and must be replaced with a new one immediately.
Sharp edges may cause serious injury. Use original Kulzer light guides only.
WARNING – Infection control:
To make treatment as safe as possible for patients and operators, the light guide and the protective light cone
must be cleaned, disinfected and sterilised before each treatment. Follow the instructions listed in paragraph 8
(Cleaning, disinfection and sterilisation) step by step.
5.5

Battery charging

CAUTION – NEW Battery: fi rst charging
The Translux Wave battery has been partially charged at the factory. Therefore the battery must be fully
charged before using the Translux Wave for the fi rst time.
NOTE – NEW Battery: fi rst charging
For the fi rst charge, this process will take about 3 hours. New batteries or batteries stored for a long time may
take longer to charge. The battery will reach its full capacity after a few full charge / discharge cycles.
2x
2x
Tabla de contenido

Solución de problemas

loading

Tabla de contenido