11
Troubleshooting
WARNING
If the below steps do not solve the problem, contact Kulzer or a local authorised Customer Service agent:
DO NOT USE the Translux Wave and DO NOT ATTEMPT to service the unit as it may cause serious injury.
If the device does not seem to be working properly, read this user manual again and then check the
following table:
PROBLEM
The charging unit does not power on.
The GREEN LED
charging unit is switched off .
GB
- 20 -
Hand piece NOT inserted into the
charging unit.
The light emission does not start
when the " STAND-BY " button
is pressed and no information
appears on the hand piece's display.
" 2 beeps " are emitted at the end of
the exposure cycle.
POSSIBLE CAUSE
The input power supply is not connected
to electrical supply outlet and / or to the
on the
charging unit power socket.
Power outlet carries no voltage.
The interchangeable adapter is not
properly inserted into the input power
supply casing.
The power cable of the input power
supply is damaged.
The input power supply is defective.
The charging unit is defective.
The charging unit contacts are short-
circuited.
Hand piece without battery.
Hand piece in " Sleep Mode ".
Empty battery.
There is not enough charge in the battery
to turn on the hand piece.
Electronics malfunction.
Defective battery.
Low battery charge.
REMEDY
Make sure that the input power supply
is securely connected to electrical
supply outlet and to the charging unit
power socket.
Use a diff erent power outlet.
Carefully read paragraph 5.2 of this user
manual.
Replace the input power supply with a
new one. (The input power supply and
its cord cannot be separated from each
other.)
Replace the input power supply with
a new one.
Contact Kulzer or local authorised
Customer Service agent.
Eliminate the cause of short circuit.
If the short circuit condition persists,
replace the charging unit contacts with
new ones.
Make reference to paragraph 8.6.2.
Insert the battery into the hand piece
(see paragraph 5.3).
Press the " STAND-BY" button to bring
the hand piece out of the " Sleep Mode ".
Make reference to paragraph 6.3.
Place the hand piece in the charging
unit and re-charge the battery
(see paragraph 5.5).
Contact Kulzer or local authorised
Customer Service agent.
To verify if the battery is defective, place
the hand piece in the charging unit.
A " prequalifi cation sequence " starts in
which Translux Wave micro-processor
checks the battery status.
This prequalifi cation sequence can last
up to max. 30 minutes.
If, during this check sequence, a " beep "
signal is emitted every second for a pe-
riod of 30 seconds and the battery icon
on the hand piece is blinking YELLOW,
the battery is detected to be defective.
Replace the battery with a new one
(see paragraph 5.3).
The " 2 beeps " signal that the residual
charge level of the battery is adequate
only for a few subsequent exposure
cycles.
Place the hand piece in the charging
unit and re-charge the battery
(see paragraph 5.5).