10 English
Bair Hugger Model 505 and Model 500/OR Temperature Management Units Operator's Manual
Specifications
Model 505
Physical Characteristics
dimensions
weight
mounting
relative noise level
hose
filtration system
recommended filter change
Temperature Characteristics
temperature control
heat generated
system time to 100°F (37.7°C)
operating temperatures
Safety System
thermostat
overcurrent protection
alarm system
certifications
classification
13 in. high x 10 in. deep x 11 in. wide
33 cm high x 25 cm deep x 28 cm wide
13.6 lb; 6.2 kg
IV pole clamp, bedrail hook with safety strap;
can be placed on hard surface
53 decibels
Detachable, flexible, washable; compatible
with the 241 fluid warming system
0.2µM level
Every 6 months or 500 hours of use.
Electronically controlled using a thermocouple sensor
1800 BTUs (average)
~17 secs
Air temperatures reaching the patient are approximately
2°C lower than the listed temperatures.
Average temperatures at the end of the hose:
HIGH:
43° ± 3°C
MED:
38° ± 3°C
LOW:
32° ± 3°C
Independent bulb and capillary
Dual fused input lines
Over-heat: flashing red light with audible alarm; heater
shuts down
IEC 60601-1; EN 60601-1-2; UL 60601-1; CAN/CSA-
C22.2, No. 601.1, EN 55011; ASTM F2196-02
Classified under IEC 60601-1 Guidelines (and other
national versions of the Guidelines) as Class I, Type BF,
Ordinary equipment, Continuous operation. Not suitable
for use in the presence of flammable anesthetic mixtures
with air or with oxygen or nitrous oxide. Classified by
Underwriters Laboratories Inc. with respect to electric
shock, fire, and mechanical hazards only, in accordance
with UL 60601-1, ASTM F2196-02, and Canadian/CSA
C22.2, No. 601.1 Classified under the Medical Device
Directive as a Class IIb device.
109.4° ± 5.4°F
100.4° ± 5.4°F
89.6° ± 5.4°F