During operation

WARNING
Danger to life and/or risk of explosion!
The following alarms indicate a danger to life and/or risk of explosion:
A2 alarm
STEL or TWA alarm
Device/channel error
► Immediately leave the hazard area.
WARNING
Incorrect measured values!
Only for diffusion mode: If water seals the gas inlets on the gas detector (e.g. in
heavy rain or if the gas detector is submerged in water), incorrect measured
values may be returned.
► With the display facing downward, shake the gas detector to remove the
water.
WARNING
Incorrect measured values!
A different reading may be shown if the gas detector experiences a considerable
impact or a significant vibration.
► When using a CatEx or IR sensor (depending on the gas detector type), a
calibration of zero-point and sensitivity must be carried out after an impact
load that results in a fresh air display not equal to zero.
CAUTION
The volume of the horn is reduced!
If water gets into the opening on the horn (e.g. by immersing the gas detector in
water or due to heavy rain), the volume of the horn can be significantly reduced.
► With the display facing downward, shake the gas detector to remove the
water.
Only using Bluetooth
in safety-critical applications. Raising the alarm on the gas detector is decisive.
Contact Dräger for a description of the API interface.
Dräger recommends limiting the usage time at temperatures below -20 °C, as
the rechargeable battery life and the display may be restricted.
®
X-am
2800
®
or API applications is insufficient for raising the alarm
|