To acknowledge an alarm:
1. Select
.
OK
4.6.3 Special state
The operation signal is deactivated during a special state. Special states are
displayed by the following visual signals:
–
Yellow LED flashing – 'warm-up 1' special state
–
Yellow LED continuously illuminated – general special state
No alarms are issued during a special state.
The special state is exited by resolving the potential error, in the event of an
intact gas detector, by switching to normal measuring mode, or automatically
after approx. 1 minute.
4.6.4 Blocking alarm
The blocking alarm protects the CatEx sensor.
If the measuring range is exceeded significantly at the CatEx channel (very high
concentration of flammable substances), a blocking alarm is triggered. This
CatEx blocking alarm can be acknowledge by switching the gas detector off and
then on again in fresh air.
For more details, refer to the technical manual.
4.7
Calling the Quick menu
The Dräger CC-Vision software can be used to save preferred features in the
Quick menu.
Possible features:
–
Fresh air calibration
–
Bump test
–
Delete peak values
–
Bluetooth
To open the Quick menu:
1. Press
3x in measuring mode.
2. Press
to scroll through the available features.
3. Press
to call the selected feature.
OK
®
X-am
2800
®
connection
|
Operation