Detectores de Gas
PointGard 2000 Serie
Dräger PointGard 2000 Serie Manuales
Manuales y guías de usuario para Dräger PointGard 2000 Serie. Tenemos
2
Dräger PointGard 2000 Serie manuales disponible para descarga gratuita en PDF: Instrucciones De Uso
Dräger PointGard 2000 Serie Instrucciones De Uso (196 páginas)
Marca:
Dräger
| Categoría:
Detectores de Gas
| Tamaño: 25.35 MB
Tabla de contenido
English
5
Tabla de Contenido
5
Safety-Related Information
7
Safety Statements
7
Operating Area and Conditions
7
Pointgard 2Xx0 Remote
7
Mechanical Installation
7
Electrical Installation
7
Commissioning
8
Calibration
8
During Operation
8
Maintenance
8
Conventions in this Document
8
Meaning of the Warnings
8
Typographical Conventions
8
Trademarks
8
Description
8
Drawing Legend
8
Feature Description
9
Pointgard 2100 EC
9
Pointgard 2200 CAT
9
Pointgard 2200 CAT Remote
9
Pointgard 2700 IR Remote
10
Intended Use
10
Approvals
10
USA and Canada
10
Mounting and Installation
10
Opening and Closing the Instrument
10
Power Supply Connector and Wiring Tables
11
Connecting the Power Supply Cable
11
Mounting the Instrument
11
Interface Connections
11
Preparation for Interface Connections
11
Relay Connections
11
4-20Ma Interface
12
Connecting the Instrument to a Controller from Dräger
12
Installing Software Dongles
12
Installation of Sensors
13
Installing EC Sensor
13
Installation of Remote Sensing Heads
13
Commissioning of the Instrument
13
Operation
13
Display, Analog Interface and Relay Status
13
Measuring Mode
13
Special States
13
Ending Special States
14
LED and Symbol Indications
14
Definitions of Indicators in the Display
15
Pointgard 2100 EC-Specific
15
Control Panel
15
Info Mode and Function Key
15
Activating Info Mode
15
Info Mode Navigation
15
Button Function
15
Using Function Key
15
Menu
15
Menu Navigation
15
Displaying Information
16
8 Calibration
16
Preparation of Calibration
16
Preparing Calibration Set-Up
16
Test Gases
16
Gas Flow for Calibrations
17
Performing Zero Calibration
17
Zero Calibration
17
Automatic Calibration
18
Performing Span Calibration
18
Span Calibration
18
Troubleshooting
18
Changing the Fuse
19
Faults
19
Warnings
19
10.3.1 Pointgard 2100 EC
20
Checking Response Time (T90)
20
Create Maintenance Plan
20
Maintenance
20
Performing a Bump Test
20
Sensor Replacement
20
10.3.2 Pointgard 2200 CAT
21
10.3.3 Pointgard 2700 IR Remote
21
Performing Display Test
21
Instrument Settings
21
Setting Date and Time
21
Setting Language
21
Setting Passwords
21
Alarm Configuration
22
Combinations of Latching and Alarm Acknowledgment
22
Configuring Function Key
22
Configuring Relay A1 or A2
22
Relays
22
Resetting to Default Settings
22
Setting Fault Relay Warm-Up
22
Switching the Alarms on or off
22
Changing the Backlight Color of the Display
23
Changing the Display Contrast
23
Changing the Display Mode
23
Configuring Alarms
23
Deactivating Dongles
23
Display Settings
23
Testing Alarms/Relays
23
4-20Ma Interface
24
Interface Settings
24
Setting Fault Current
24
Setting the DIP Switch Pins
24
Switching the Warning Signal on or off
24
Setting Analog Offset
25
Setting Analog Span
25
Setting Maintenance Signal
25
Setting Static Maintenance Current
25
Setting Warning Current
25
Setting Warning Interval
25
Testing the Analog Interface
25
Beam Block
26
Capture Range
26
Resetting Sensor
26
Configuration of Integrated Alarm Devices
24
DIP Switch
24
Full Scale Deflection
24
Information Regarding the Warning Signal
24
Sensor Settings Pointgard 2Xx0
26
Setting Calibration Interval
26
Setting Capture Range
26
Setting Sensor Lock
26
Switching Automatic Calibration on or off
26
Switching Capture Range off
26
Deactivating Dongles
27
Gas Settings
27
Sensor Settings Pointgard 2100 EC
27
Sensor Settings Pointgard 2200 CAT
27
Sensor Test
27
Sensor Type
27
Setting DQ Sensor Latching
27
Software Dongles for EC Sensors
27
Factory Default Settings
28
Gas Settings
28
Pointgard 2100 EC
28
Pointgard 2Xx0
28
Sensor Settings Pointgard 27X0 IR
28
Sensor-Specific Values
28
Settings Which Can be Changed Via the Menu
28
Pointgard 2200 CAT Drägersensor ® DQ
29
Pointgard 2200 CAT Drägersensor ® LC
29
Pointgard 2200 CAT Remote DSIR
29
Pointgard 2700 IR
29
Pointgard 2720 IR
30
AC Version
31
Disposal
31
Measuring Ranges
31
Power Supply
31
Signal Transmission to Control Unit
31
Technical Data
31
Cable Specifications Catex Remote
32
Cable Specifications PIR Remote
32
DC Version
32
Environmental Parameters
32
Physical Specifications
32
Tightening Torque for Instrument Threads
32
Accessories and Spare Parts
33
Pointgard 2Xx0 Accessories
33
Pointgard 2Xx0 Series
33
Sensor Range and Default Alarm Values
33
Sensor Range Catalytic Bead (Catex) Sensors
33
Sensor Range Electrochemical (EC) Sensors
33
Spare Parts
33
Mounting Screws
34
Pointgard 2100 EC
34
Pointgard 2100 EC Accessories
34
Pointgard 2200 CAT
34
Pointgard 2200 CAT Accessories
34
Pointgard 27X0 IR
34
Power Supply Cables
34
Sensors
34
Deutsch
35
Sicherheitsbezogene Informationen
37
Einsatzbereich und Einsatzbedingungen
37
Elektrische Installation
37
Mechanische Installation
37
Pointgard 2Xx0 Remote
37
Sicherheitshinweise
37
Bedeutung der Warnhinweise
38
Inbetriebnahme
38
Justierung
38
Konventionen in diesem Dokument
38
Typografische Konventionen
38
Wartung
38
Während des Betriebs
38
Beschreibung
39
Funktionsbeschreibung
39
Marken
39
Pointgard 2100 EC
39
Zeichnungslegende
39
Montage und Installation
40
Pointgard 2200 CAT
40
Pointgard 2200 CAT Remote
40
Pointgard 2700 IR Remote
40
Verwendungszweck
40
Öffnen und Schließen des Geräts
40
Anschluss der Schnittstelle Vorbereiten
41
Montage des Geräts
41
Relaisverbindungen
41
Schnittstellenverbindungen
41
Steckverbinder für Spannungsversorgung und Verdrahtungstabellen
41
Verbinden des Netzanschlusskabels
41
4-20-Ma-Schnittstelle
42
Anzeige, Analogschnittstelle und Relaisstatus
43
Inbetriebnahme des Geräts
43
Installation von Abgesetzten Messköpfen
43
Betrieb
43
Das Gerät an ein Dräger-Auswertegerät Anschließen
43
EC-Sensor Installieren
43
Installation von Sensoren
43
Software-Dongles Installieren
43
Beendigung der Sonderzustände
44
Messbetrieb
44
Sonderzustände
44
Bedienfeld
45
Erklärung der Symbole auf dem Display
45
Leds und Symbolanzeigen
45
Pointgard 2100 EC-Spezifisch
45
Aktivieren des Info-Modus
46
Darstellung von Informationen
46
Funktionstaste Verwenden
46
Info-Modus und Funktionstaste
46
Kennwörter
46
Menü
46
Navigation IM Infomodus
46
Navigation IM Menü
46
Zugriff auf Menüs
46
Einrichten der Justierung
47
Justierung
47
Justierung Vorbereiten
47
Prüfgase
47
Gasdurchfluss für Justierungen
48
Nullpunktjustierung
48
Nullpunktjustierung Durchführen
48
Austausch der Sicherung
49
Automatische Kalibrierung
49
Empfindlichkeitsjustierung
49
Empfindlichkeitsjustierung Durchführen
49
Störungsbeseitigung
49
Fehler
50
Warnungen
50
Ansprechzeit Prüfen (T90)
51
Begasungstest Durchführen
51
Instandhaltung
51
Sensoren Austauschen
51
Wartungsplan Erstellen
51
10.3.1 Pointgard 2100 EC
52
10.3.2 Pointgard 2200 CAT
52
10.3.3 Pointgard 2700 IR Remote
52
Display-Test Durchführen
52
Geräteeinstellungen
52
Kennwörter Einstellen
52
Auf Standardeinstellungen Zurücksetzen
53
Datum und Uhrzeit Einstellen
53
Fehlerrelais Einlaufen 1 Einstellen
53
Funktionstaste Konfigurieren
53
Kombinationen von Selbsthaltenden Alarmen und Alarmquittierung
53
Relais A1 oder A2 Konfigurieren
53
Sprache Einstellen
53
Alarme Ein- oder Ausschalten
54
Alarme/Relais Testen
54
Alarmkonfiguration
54
Konfigurieren von Alarmen
54
Anzeigemodus Ändern
55
DIP-Schalter
55
Displayeinstellungen
55
Dongles Deaktivieren
55
Einstellen der Schiebehebel des DIP-Schalters
55
Farbe der Hintergrundbeleuchtung des Displays Ändern
55
Konfiguration der Integrierten Alarmgeber
55
Kontrast des Displays Ändern
55
4-20-Ma-Schnittstelle
56
Einstellung des Statischen Wartungsstroms
56
Einstellung des Wartungssignals
56
Fehlerstrom Einstellen
56
Informationen zum Warnsignal
56
Messbereichsendwert
56
Schnittstelleneinstellungen
56
Warnungsintervall Einstellen
56
Warnungssignal Ein- oder Ausschalten
56
Warnungsstrom Einstellen
56
Analogempfindlichkeit Einstellen
57
Analogschnittstelle Testen
57
Beam Block
57
Automatische Justierung Ein- oder Ausschalten
58
Fangbereich
58
Fangbereich Deaktivieren
58
Fangbereich Einstellen
58
Justierintervall Einstellen
58
Sensor Zurücksetzen
58
Sensor-Sperrfunktion Einstellen
58
Sensoreinstellungen Pointgard 2Xx0
58
Dongles Deaktivieren
59
Gaseinstellungen
59
Sensoreinstellungen Pointgard 2100 EC
59
Sensoreinstellungen Pointgard 2200 CAT
59
Sensortest
59
Sensortyp
59
Software-Dongle für EC-Sensoren
59
Einstellen der DQ-Sensorverriegelung
60
Einstellungen, die über das Menü Geändert werden können
60
Gaseinstellungen
60
Pointgard 2Xx0
60
Sensoreinstellungen Pointgard 27X0 IR
60
Werkseinstellungen
60
Pointgard 2100 EC
61
Pointgard 2200 CAT Drägersensor ® DQ
61
Pointgard 2200 CAT Drägersensor ® LC
61
Pointgard 2200 CAT Remote DSIR
61
Sensorspezifische Werte
61
Entsorgung
63
Messbereiche
63
Signalübertragung zur Auswerteeinheit
63
Technische Daten
63
AC-Version
64
DC-Version
64
Physische Spezifikationen
64
Spannungsversorgung
64
Umgebungsparameter
64
Anzugsdrehmoment für Gerätegewinde
65
Kabelspezifikationen Abgesetzter Catex-Sensor
65
Kabelspezifikationen Abgesetzter PIR-Sensor
65
Sensorbereich - Wärmetönungssensoren (Catex)
65
Sensorbereich Elektrochemische (EC) Sensoren
65
Sensorbereich und Standard-Alarmwerte
65
Befestigungsschrauben
66
Ersatzteile
66
Netzanschlusskabel
66
Pointgard 2100 EC
66
Serie Pointgard 2Xx0
66
Zubehör für Pointgard 2100 EC
66
Zubehör für Pointgard 2Xx0
66
Zubehör und Ersatzteile
66
Pointgard 2200 CAT
67
Pointgard 27X0 IR
67
Sensoren
67
Zubehör für Pointgard 2200 CAT
67
Français
68
Informations Relatives à la Sécurité
70
Consignes de Sécurité
70
Installation Mécanique
70
Installation Électrique
70
Pointgard 2Xx0 à Distance
70
Zone et Conditions D'utilisation
70
Calibrage
71
Conventions Typographiques
71
Conventions Utilisées Dans Ce Document
71
Maintenance
71
Mise en Service
71
Pendant le Fonctionnement
71
Signification des Avertissements
71
Description
72
Description des Fonctions
72
Légende du Schéma
72
Marques
72
Pointgard 2100 EC
72
Domaine D'application
73
Homologations
73
Pointgard 2200 CAT
73
Pointgard 2200 CAT Remote
73
Pointgard 2700 IR Remote
73
États-Unis et Canada
73
Branchement du Câble D'alimentation
74
Connecteur D'alimentation et Tableaux de Câblage
74
Montage et Installation
74
Ouvrir et Fermer L'instrument
74
Connexions à Interface
75
Montage de L'instrument
75
Préparation des Connexions à Interface
75
Raccords de Relais
75
Connexion de L'équipement à un Système de Commande Dräger
76
Installation de Modules de Tests Complémentaires
76
Installation des Capteurs
76
Installation des Capteurs à Distance
76
Installer le Capteur EC
76
Affichage, Interface Analogique et État du Relais
77
Fonctionnement
77
Mise en Service de L'appareil
77
Mode de Mesure
77
Statuts Spéciaux
77
Indications des LED et des Symboles
78
Terminer les Status Spéciaux
78
Accéder au Menu
79
Activer le Mode Info
79
Définitions des Indicateurs Affichés à L'écran
79
Mode Info et Touche de Fonction
79
Mots de Passe
79
Navigation Dans le Menu
79
Navigation en Mode Info
79
Panneau de Commande
79
Pointgard 2100 Spécifique EC
79
Utiliser la Touche de Fonction
79
Affichage D'informations
80
Gaz Étalon
80
Débit de Gaz pour Calibrages
81
Préparation au Calibrage
81
Préparer le Réglage du Calibrage
81
Réglage du Zéro
81
Calibrage Sensibilité
82
Effectuer le Réglage du Point Zéro
82
Réalisation du Calibrage Sensibilité
82
Calibrage Automatique
83
Dépannage
83
Remplacer le Fusible
83
Rapport D'info
84
Réalisation D'un Test au Gaz
84
10.3.1 Pointgard 2100 EC
85
10.3.2 Pointgard 2200 CAT
85
Calendrier de Maintenance
85
Contrôle du Temps de Réponse (T90)
85
Remplacement du Capteur
85
10.3.3 Pointgard 2700 IR Remote
86
Configuration des Mots de Passe
86
Configurer la Touche de Fonction
86
Effectuer un Test D'affichage
86
Réglages de L'appareil
86
Régler la Date et L'heure
86
Régler la Langue
86
Réinitialisation des Réglages Par Défaut
86
Activation ou Désactivation des Alarmes
87
Associations D'auto-Maintien et D'acquittement D'alarme
87
Configuration de L'alarme
87
Configuration des Alarmes
87
Configuration du Relais A1 ou A2
87
Réglage Relais Défaut Période de Stabilisation
87
Changer la Couleur du Rétro-Éclairage de L'affichage
88
Changer le Mode D'affichage
88
Configuration des Dispositifs D'alarme Intégrés
88
Désactiver les Dongles
88
Modifier le Contraste de L'affichage
88
Réglages de L'affichage
88
Test des Alarmes / Relais
88
Commutateur DIP
89
Dépassement D'échelle
89
Informations Relatives au Signal D'avertissement
89
Interface 4-20Ma
89
Pour Activer ou Désactiver le Courant D'avertissement
89
Réglage des Broches du Commutateur DIP
89
Réglage du Signal D'erreur
89
Réglages de L'interface
89
Ajustage Signal Sortie
90
Effectuer un Test Avec L'interface Analogique
90
Réglage de L'intervalle D'avertissement
90
Réglage du Courant de Maintenance Statique
90
Réglage du Signal de Maintenance
90
Régler le Courant D'avertissement
90
Activation ou Désactivation du Calibrage Automatique
91
Beam Block
91
Désactivation de la Plage de Capture
91
Plage de Capture
91
Réglage de la Plage de Capture
91
Réglages Capteur Pointgard 2Xx0
91
Réinitialiser le Capteur
91
Configuration de Gaz
92
Configuration Période de Calibrage
92
Dongles Logciels pour Capteurs EC
92
Désactiver les Dongles
92
Réglage du Verrouillage du Capteur
92
Réglages Capteur Pointgard 2100 EC
92
Réglages Capteur Pointgard 2200 CAT
92
Test du Capteur
92
Type de Capteur
92
Configuration de Gaz
93
Pointgard 2Xx0
93
Réglage du Verrouillage du Capteur DQ
93
Réglages du Capteur Pointgard 27X0 IR
93
Réglages Modifiables à L'aide du Menu
93
Réglages Par Défaut D'usine
93
Pointgard 2100 EC
94
Pointgard 2200 CAT Drägersensor ® DQ
94
Valeurs Spécifiques du Capteur
94
Pointgard 2200 CAT Drägersensor ® LC
95
Pointgard 2200 CAT Remote DSIR
95
Pointgard 2700 IR
95
Elimination
96
Pointgard 2720 IR
96
Alimentation
97
Caractéristiques Techniques
97
Plages de Mesure
97
Transmission du Signal Vers L'unité de Commande
97
Version AC
97
Version DC
97
Couple de Serrage pour les Alésages de L'instrument
98
Paramètres Environnementaux
98
Spécifications de Câbles pour le Capteur PIR à Distance
98
Spécifications du Câble pour Capteur Catex à Distance
98
Spécifications Physiques
98
Accessoires et Pièces Détachées
99
Accessoires Pointgard 2Xx0
99
Pièces Détachées
99
Plage des Capteurs à Perle Catalytique (Catex)
99
Plage des Capteurs Électrochimiques (EC)
99
Portée du Capteur et Valeurs D'alarme Par Défaut
99
Série Pointgard 2Xx0
99
Accessoires Pointgard 2100 EC
100
Accessoires Pointgard 2200 CAT
100
Capteurs
100
Câbles D'alimentation
100
Pointgard 2100 EC
100
Pointgard 2200 CAT
100
Pointgard 27X0 IR
100
Vis de Montage
100
Dräger PointGard 2000 Serie Instrucciones De Uso (190 páginas)
Marca:
Dräger
| Categoría:
Instrumentos de Medición
| Tamaño: 6.12 MB
Tabla de contenido
English
5
Tabla de Contenido
5
Safety-Related Information
7
Safety Statements
7
Operating Area and Conditions
7
Pointgard 2200 CAT Remote
7
Mechanical Installation
7
Electrical Installation
7
Commissioning
8
Calibration
8
During Operation
8
Maintenance
8
Conventions in this Document
8
Meaning of the Warning Notes
8
Trademarks
8
Description
8
Product Overview
8
Caption
9
Feature Description
9
Pointgard 2100 EC
9
Pointgard 2200 CAT
9
Intended Use
10
Approvals
10
USA and Canada
10
Mounting and Installation
10
Opening and Closing the Instrument
10
Preparing the Power Supply Cable
10
Power Supply Connector and Wiring Tables
11
Connecting the Power Supply Cable
11
Mounting the Instrument
11
Interface Connections
11
Preparation for Interface Connections
11
Relay Connections
11
4-20Ma Interface
12
Connecting the Instrument to a Controller from Dräger
12
Installing Software Dongles
12
Installation of Sensors
13
Installation of EC Sensors
13
Installing EC Sensors
13
Installation of the EC Remote Sensor Enclosure
13
Connection of the Catex Remote Version
13
Connecting the Catex Remote Version
14
Commissioning of the Instrument
14
Operation
14
Display. Analog Interface and Relay Status
14
Measuring Mode
15
Special States
15
Ending Special States
15
LED and Symbol Indications
16
Definitions of Indicators in the Display
16
Pointgard 2100 EC-Specific
16
Control Panel
16
Button Function
16
Info Mode and Function Key
16
Activating Info Mode
16
Info Mode Navigation
16
Using Function Key
16
Menu
16
Passwords
17
Entering the Menu
17
Displaying Information
17
8 Calibration
17
Test Gases
17
Preparation of Calibration
17
Preparing Calibration Set-Up
18
Gas Flow for Calibrations
18
Zero Calibration
18
Performing Zero Calibration
18
Span Calibration
19
Performing Span Calibration
19
Automatic Calibration
19
Performing Automatic Calibration
19
Troubleshooting
20
Changing the Fuse
20
Faults
20
Warnings
20
Maintenance
21
Performing a Bump Test
21
Create Maintenance Plan
21
Sensor Replacement
21
10.3.1 Pointgard 2100 EC
21
10.3.2 Pointgard 2200 CAT
22
Instrument Settings
22
Alarm Configuration
22
Switching the Alarms on or off
22
Configuring Alarms
22
Relays
23
Configuring Relay A1 or A2
23
Setting Fault Relay Warm-Up
23
Testing Alarms/Relays
23
Setting Passwords
23
Setting Date and Time
23
Setting Language
23
Configuring Function Key
23
Resetting to Default Settings
24
Deactivating Dongles
24
Data-Logger Settings
24
Information Regarding the Data-Logger
24
Displaying Graph
24
Switching Data-Logger on or off
24
Setting the Sampling Time
24
Setting Peak/Average
24
Switching Triggering Mode on or off
24
Setting Trigger Value
24
Clearing Data-Logger
25
Display Settings
25
Changing the Backlight Color of the Display
25
Changing the Display Contrast
25
Changing the Display Mode
25
Performing Display Test
25
Configuration of Integrated Alarm Devices
25
DIP Switch
25
Setting the DIP Switch Pins
25
Green/Red Strobe Combination
26
Default Setting: Horn Active on A1 Alarms and on A2 Alarms
26
Setting: Horn Only Active on A2 Alarms
26
Setting the Horn to Only Active on A2 Alarms
26
Amber/Red Strobe Combination
27
Default Setting: Horn Only Active on A2 Alarms
27
Setting: Horn Only Active on A1 Alarms
27
Setting: Horn Active on A1 Alarms and on A2
27
Acknowledging Horn Alarms (Horn Silence)
27
Interface Settings
28
4-20Ma Interface
28
Full Scale Deflection
28
Setting Fault Current
28
Information Regarding the Warning Signal
28
Switching the Warning Signal on or off
28
Setting Warning Interval
28
Setting Warning Current
28
Setting Maintenance Signal
28
Setting Static Maintenance Current
28
Testing the Analog Interface
29
Sensor Settings Pointgard 2X00
29
Switching Automatic Calibration on or off
29
Setting Capture Range
29
Resetting Sensor
29
Setting Calibration Interval
29
Sensor Settings Pointgard 2100 EC
29
Sensor Test
29
Gas Settings
30
Setting Sensor Lock
30
Sensor Settings Pointgard 2200 CAT
30
Sensor Type
30
Setting DD Sensor Latching
30
Disposal
30
Technical Data
31
Signal Transmission to Control Unit
31
Power Supply
31
AC Version
31
DC Version
31
Physical Specifications
31
Environmental Parameters
31
Tightening Torque for Instrument Threads
32
Cable Specifications Catex Remote
32
Sensor Range and Default Alarm Values
32
Sensor Range Catalytic Bead (Catex) Sensors
32
Sensor Range Electrochemical (EC) Sensors
32
Accessories and Spare Parts
32
Pointgard 2X00 Accessories
33
Pointgard 2100 EC Accessories
33
Pointgard 2200 CAT Accessories
33
Power Supply Cables
33
Mounting Screws
33
Spare Parts
33
Pointgard 2000 Versions and Part Numbers
33
Deutsch
34
Sicherheitsbezogene Informationen
36
Sicherheitshinweise
36
Einsatzbereich und Betriebsbedingungen
36
Pointgard 2200 CAT Remote
36
Mechanische Installation
36
Elektrische Installation
36
Inbetriebnahme
37
Justierung
37
Während des Betriebs
37
Wartung
37
Konventionen in diesem Dokument
37
Bedeutung der Warnhinweise
37
Markenzeichen
38
Beschreibung
38
Produktübersicht
38
Legende
38
Funktionsbeschreibung
38
Pointgard 2100 EC
38
Pointgard 2200 CAT
38
Verwendungszweck
39
Montage und Installation
39
Öffnen und Schließen des Geräts
39
Vorbereiten des Netzanschlusskabels
39
Steckverbinder für Spannungsversorgung und Verdrahtungstabellen
39
Verbinden des Netzanschlusskabels
40
Montage des Geräts
40
Schnittstellenverbindungen
40
Anschluss der Schnittstelle Vorbereiten
40
Relaisverbindungen
40
4-20-Ma-Schnittstelle
41
Das Gerät an ein Dräger-Auswertegerät Anschließen
41
Software-Dongles Installieren
42
Installation von Sensoren
42
Installation von EC-Sensoren
42
EC-Sensoren Installieren
42
Installation des Gehäuses für Abgesetzte EC-Sensoren
42
Anschließen der Catex-Remote-Version
43
Inbetriebnahme des Geräts
44
Betrieb
44
Anzeige - Analogschnittstelle und Relais-Zustand
44
Messbetrieb
44
Sonderzustände
44
Beendigung der Sonderzustände
45
Leds und Symbolanzeigen
45
Erklärung der Symbole auf dem Display
46
Pointgard 2100 EC-Spezifisch
46
Bedienfeld
46
Info-Modus und Funktionstaste
46
Aktivieren des Info-Modus
46
Navigation IM Infomodus
46
Funktionstaste Verwenden
46
Menü
46
Navigation IM Menü
46
Kennwörter
46
Darstellung von Informationen
47
Justierung
47
Prüfgase
47
Kalibrierung Vorbereiten
47
Einrichten der Justierung
48
Gasdurchfluss für Justierungen
48
Nullpunktjustierung
48
Nullpunktjustierung Durchführen
48
Empfindlichkeitsjustierung
49
Empfindlichkeitsjustierung Durchführen
49
Automatische Justierung
49
Automatische Justierung Durchführen
49
Störungsbeseitigung
50
Austausch der Sicherung
50
Fehler
50
Warnungen
51
Instandhaltung
51
Begasungstest Durchführen
51
Wartungsplan Erstellen
51
Sensoren Austauschen
51
10.3.1 Pointgard 2100 EC
51
10.3.2 Pointgard 2200 CAT
52
Geräteeinstellungen
52
Alarmkonfiguration
52
Alarme Ein- oder Ausschalten
52
Alarme Konfigurieren
52
Relais
53
Relais A1 oder A2 Konfigurieren
53
Fehlerrelais Einlaufen 1 Einstellen
53
Alarme/Relais Testen
53
Kennwörter Einstellen
54
Datum und Uhrzeit Einstellen
54
Sprache Einstellen
54
Funktionstaste Konfigurieren
54
Auf Standardeinstellungen Zurücksetzen
54
Dongles Deaktivieren
54
Datenlogger-Einstellungen
54
Informationen über den Datenlogger
54
Grafik Anzeigen
54
Datenlogger Ein- oder Ausschalten
54
Auslösemodus Ein- oder Ausschalten
55
Triggerschwelle Einstellen
55
Stack/Roll-Funktion Einstellen
55
Datenlogger Löschen
55
Displayeinstellungen
55
Farbe der Hintergrundbeleuchtung des Displays Ändern
55
Kontrast des Displays Ändern
55
Anzeigemodus Ändern
55
Display-Test Durchführen
55
Konfiguration der Integrierten Alarmgeber
56
DIP-Schalter
56
Einstellen der Schiebehebel des DIP-Schalters
56
Kombination aus Grüner/Roter Stroboskopleuchte
56
Standardeinstellung: Signalhorn Aktiv bei A1- und A2-Alarmen
56
Einstellung: Signalhorn ist nur bei A2-Alarmen Aktiv
57
Einstellen des Signalhorns, damit es nur bei A2-Alarmen Aktiv ist
57
Kombination aus Bernsteinfarbener/Roter Stroboskopleuchte
57
Standardeinstellung: Signalhorn ist nur bei A2-Alarmen Aktiv
57
Einstellung: Signalhorn ist nur bei A1-Alarmen Aktiv
57
Einstellung: Signalhorn Aktiv bei A1- und A2-Alarmen
58
Quittieren von Signalhorn-Alarmen (Signalhorn Stummschalten)
58
Schnittstelleneinstellungen
58
4-20-Ma-Schnittstelle
58
Messbereichsendwert
58
Fehlerstrom Einstellen
58
Informationen zum Warnungssignal
59
Warnungssignal Ein- oder Ausschalten
59
Warnungsintervall Einstellen
59
Warnungsstrom Einstellen
59
Wartungssignal Einstellen
59
Statischen Wartungsstrom Einstellen
59
Analogen Offset Einstellen
59
Analogempfindlichkeit Einstellen
59
Analogschnittstelle Testen
59
Sensoreinstellungen Pointgard 2X00
60
Automatische Justierung Ein- oder Ausschalten
60
Fangbereich Einstellen
60
Sensor Zurücksetzen
60
Justierintervall Einstellen
60
Sensoreinstellungen Pointgard 2100 EC
60
Sensorselbsttest
60
Gaseinstellungen
60
Sensor-Sperrfunktion Einstellen
61
Sensoreinstellungen Pointgard 2200 CAT
61
Sensortyp
61
Gaseinstellungen
61
Einstellen der DD-Sensorverriegelung
61
Entsorgung
61
Technische Daten
62
Signalübertragung zur Auswerteeinheit
62
Spannungsversorgung
62
AC-Version
62
DC-Version
62
Physische Spezifikationen
62
Umgebungsparameter
62
Anzugsdrehmoment für Gerätegewinde
63
Kabelspezifikationen Abgesetzter Catex-Sensor
63
16.7 Sensorbereich und Standard-Alarmwerte
63
Sensorbereich - Wärmetönungssensoren (Catex)
63
Sensorbereich - Elektrochemische Sensoren (EC-Sensoren)
63
Zubehör und Ersatzteile
64
Zubehör für Pointgard 2X00
64
Zubehör für Pointgard 2100 EC
64
Zubehör für Pointgard 2200 CAT
64
Netzanschlusskabel
64
Befestigungsschrauben
64
Ersatzteile
64
Versionen und Sachnummern Pointgard 2000
65
Français
66
Informations Relatives à la Sécurité
68
Consignes de Sécurité
68
Zone et Conditions D'utilisation Prévues
68
Pointgard 2200 CAT à Distance
68
Installation Mécanique
68
Installation Électrique
68
Mise en Service
69
Calibrage
69
Pendant le Fonctionnement
69
Maintenance
69
Conventions Utilisées Dans Ce Document
69
Signification des Avertissements
69
Marques Déposées
70
Description
70
Vue D'ensemble du Produit
70
Légende
70
Description des Fonctions
70
Pointgard 2100 EC
70
Pointgard 2200 CAT
71
Domaine D'application
71
Homologations
71
États-Unis et Canada
71
Montage et Installation
71
Ouvrir et Fermer L'instrument
71
Préparation du Câble D'alimentation
72
Connecteur D'alimentation et Tableaux de Câblage
72
Branchement du Câble D'alimentation
72
Montage de L'instrument
72
Connexions à Interface
72
Préparation des Connexions à Interface
72
Raccords de Relais
73
Interface 4-20Ma
73
Connexion de L'équipement au Système de Surveillance Dräger
74
Installation de Dongles Logiciels
74
Installation des Capteurs
74
Installation des Capteurs EC
74
Installer les Capteurs EC
74
Installation du Boîtier pour Capteur EC à Distance
74
Connexion de la Version Catex à Distance
75
Connecter la Version Catex à Distance
76
Fonctionnement
76
Affichage, Interface Analogique et État du Relais
76
Mode de Mesure
77
Statuts Spéciaux
77
Terminer les Status Spéciaux
77
Indications des LED et des Symboles
78
Définitions des Indicateurs Affichés à L'écran
78
Pointgard 2100 Spécifique EC
78
Panneau de Commande
78
Mode Info et Touche de Fonction
78
Activer le Mode Info
78
Navigation en Mode Info
78
Utiliser la Touche de Fonction
79
Navigation Dans le Menu
79
Mots de Passe
79
Accéder au Menu
79
69 7.6.4 Affichage des Informations
79
Gaz Étalons
80
Préparation au Calibrage
80
Préparer le Réglage du Calibrage
80
Débit de Gaz pour Calibrages
80
Réglage du Zéro
80
Effectuer le Réglage du Point Zéro
81
Calibrage Sensibilité
81
Réalisation du Calibrage Sensibilité
81
Calibrage Automatique
82
Réalisation du Calibrage Automatique
82
Mise en Service de L'appareil
76
Dépannage
82
Remplacer le Fusible
82
Rapport D'info
83
Effectuer un Test au Gaz
84
Calendrier de Maintenance
84
Remplacement du Capteur
84
10.3.1 Pointgard 2100 EC
84
10.3.2 Pointgard 2200 CAT
84
Réglages de L'appareil
85
Configuration de L'alarme
85
Activation ou Désactivation des Alarmes
85
Configurer les Alarmes
85
Relais
85
Configuration du Relais A1 ou A2
85
Réglage Relais Défaut Période de Stabilisation
85
Test des Alarmes / Relais
86
Configuration des Mots de Passe
86
Régler la Date et L'heure
86
Régler la Langue
86
Configurer la Touche de Fonction
86
Réinitialisation des Réglages Par Défaut
86
Désactiver les Dongles
86
Configuration Data-Logger
87
Information Data-Logger
87
Afficher le Graphe
87
Mémoire Activée ou Désactivée
87
Régler la Période D'échantillonnage
87
Réglage de la Valeur de Pic / Moyenne
87
Activation ou Désactivation du Mode de Déclenchement
87
Régler la Valeur de Déclenchement
87
Régler la Fonction Empilement / Écrasement
87
Effacer la Mémoire
87
Modifier le Contraste de L'affichage
88
Changer le Mode D'affichage
88
Effectuer un Test D'affichage
88
Configuration des Dispositifs D'alarme Intégrés
88
Commutateur DIP
88
Réglage des Broches du Commutateur DIP
88
Combinaison Clignotants Vert et Rouge
89
Réglage Par Défaut : L'avertisseur Sonore Est Activé pour les Alarmes A1 et les Alarmes A2
89
Réglage : L'avertisseur Sonore Est Activé Uniquement en cas D'alarmes A2
89
Régler L'avertisseur Sonore pour Qu'il Ne S'active Qu'en cas D'alarmes A2
89
Combinaison Clignotants Orange et Rouge
90
Réglage Par Défaut : L'avertisseur Sonore Est Activé Uniquement en cas D'alarmes A2
90
Réglage : L'avertisseur Sonore Est Activé Uniquement en cas D'alarmes A1
90
Réglage : L'avertisseur Sonore Est Activé pour les Alarmes A1 et les Alarmes A2
90
Acquitter L'avertisseur Sonore (Arrêter L'avertisseur Sonore)
90
Réglages de L'interface
91
Interface 4-20Ma
91
Dépassement D'échelle
91
Réglage du Signal D'erreur
91
Information Concernant le Signal D'avertissement
91
Pour Activer ou Désactiver le Courant D'avertissement
91
Réglage de L'intervalle D'avertissement
91
Régler le Courant D'avertissement
91
Réglage du Signal D'avertissement
91
Régler le Courant du Signal de Maintenance
91
Ajustage Signal Sortie
92
Effectuer un Test Avec L'interface Analogique
92
Réglages Capteur Pointgard 2X00
92
Activation ou Désactivation du Calibrage Automatique
92
Réglage de la Plage de Capture
92
Réinitialiser le Capteur
92
Configuration Période de Calibrage
93
Réglages Capteur Pointgard 2100 EC
93
Test Capteur
93
Configuration de Gaz
93
Réglage de la Fonction Verrouillage Capteur
93
Réglages Capteur Pointgard 2200 CAT
93
Type de Capteur
93
Réglage Auto-Maintien du Capteur DD
94
Caractéristiques Techniques
94
Transmission du Signal Vers L'unité de Commande
94
Version AC
94
Version DC
95
Spécifications Physiques
95
Paramètres Environnementaux
95
Couple de Serrage pour les Alésages de L'instrument
95
Spécifications du Câble pour Capteur Catex à Distance
95
Échelle Capteur et Alarme Par Défaut
95
Plage des Capteurs à Perle Catalytique (Catex)
95
Plage des Capteurs Électrochimiques (EC)
96
Accessoires et Pièces Détachées
96
Accessoires Pointgard 2X00
96
Accessoires Pointgard 2100 EC
96
Accessoires Pointgard 2200 CAT
96
Câbles D'alimentation
97
16.8.5 Vis de Montage
97
Pièces Détachées
97
Versions et Références de Pièces du
97
Productos relacionados
Dräger PointGard 2100 EC
Dräger PointGard 2200 CAT Remote
Dräger PointGard 27 0 IR Remote
Dräger Oxylog 2000 plus
Dräger Polaris 200
Dräger 200 ECO
Dräger VarioGard 2320
Dräger VarioGard 2300
Dräger Inhalog 2 plus
Dräger REGARD 2410
Dräger Categorias
Equipo Medico
Equipo Respiratorio
Instrumentos de Medición
Equipo de Pruebas
Detectores de Gas
Más Dräger manuales