C3.4 Closing the Chamber Door
The chamber door utilises a two stage closure mechanism. Carefully close the chamber door to
•
the first click and do not force it shut. The door will automatically close further to seal and lock at
the start of the HLD cycle.
If the door is not properly closed, a screen message will prompt you to close the chamber door.
•
DO NOT force open or manually unlock the chamber door during a HLD cycle.
C3.5 Disinfecting the Probe
Once the door is closed, the trophon2 device will request confirmation that the probe is clean and dry
prior to HLD. The door will lock upon confirmation.
1.
Scan the trophon AcuTrace Operator Card when prompted (if AcuTrace is enabled) and follow
the onscreen instructions.
The screen will ask:
Is the probe clean and dry?
2.
If the probe has NOT been cleaned and dried, select No. Remove the probe, clean and dry the
probe in accordance with C3.1 Preparing the Probe and repeat from C3.3 Positioning the Probe.
If the probe is clean proceed with the step below.
Figure 14
Figure 15