8 5 Tap-On™ Technology Wireless Foot
Pedal Operation
Using Foot Pedal in Tap-On
For scaling procedures,Tap-On
need to hold the pedal down. Tapping the foot pedal once
activates ultrasonic power or rinse mode for approximately
4 minutes. Tapping the foot pedal while in Tap-On
disables the ultrasonic power and water flow. Boost is still
available while scaling in Tap-On
depress the foot pedal to the second position (all the way to
the floor) to activate and hold as long as boost is desired.
Release foot pedal to return to Tap-On
DEPRESSED
NON-DEPRESSED
1st POSITION
TIPS:
Tap-On
Technology will not run water unless an
TM
insert is in the handpiece
A sensor in the handpiece holder will prevent
Tap-On
Technology from operating when the
TM
handpiece is in the holder.
If the foot pedal is not tapped quickly, it will
function in a conventional manner.
How to Disable and Enable Tap-On
Tap-On
Technology feature can be disabled by
TM
simultaneously holding the Purge and Turbo buttons for a
period of approximately 5 seconds. The two buttons will blink
approximately 6 times. When the buttons are released, they
will blink an additional 6 times to confirm Tap-On
disabled.
Tap-On
Technology feature can be enabled by
TM
simultaneously holding the Purge and Turbo buttons for a
period of approximately 5 seconds. The two buttons will blink
approximately 6 times to confirm Tap-On
Using Foot Pedal without Tap-On
For scaling operation, the first position activates both the
ultrasonic energy and lavage at the insert tip. The second
position activates the Boost Mode. The Boost Mode (fully
depressed Tap-On
Foot Pedal) increases the ultrasonic
TM
power level for quick removal of tenacious deposits without
adjusting the power level knob. To deactivate Boost Mode,
release Tap-On
Foot Pedal to first position.
TM
8 6 Accessories and User
Replaceable Parts
8 6 1 Accessories
1. AC Power Cord
2. Tap-On
Technology Wireless Foot Pedal
TM
3. Auxiliary Tap-On
Foot Pedal Power Cable
TM
®
4. Cavitron Steri- MATE
Sterilizable Handpiece
5. Cavitron 30K Ultrasonic Inserts
Mode
T M
Technology eliminates the
TM
mode
TM
mode. To use boost, simply
TM
mode.
T M
DEPRESSED
2nd POSITION
Technology
TM
has been
TM
TM
has been enabled.
Mode
T M
6. Cavitron DualSelect Dispensing system
®
7. Steri- MATE
Grip
®
8. Steri- MATE
Light Lighted Sleeve
8 6 2 User Replaceable Part Kits
1.
Cavitron Insert Replacement O-ring Kits, 12/packs
Part Number 62351 (black) for plastic and soft grips
Part Number 62605 (green) for metal grips
2.
Handpiece Cable O-Ring, Part Number 79357
3.
Lavage (Water) Filter, 10/Pack, Part Number 90158
For detailed information, contact your local DENTSPLY
Professional Representative or authorized DENTSPLY
Professional Distributor.
SECTION 9: System Setup,
Operation and Techniques
for Use
9 1 Handpiece Setup
9 1 Handpiece Setup
•
This handpiece is sterilizable. Refer to Infection Control
Information Booklet for sterilization instructions prior to
using handpiece.
•
Connect the Handpiece to the Cable Assembly by
aligning the electrical connections. If Cable Assembly
does not seat into the handpiece, gently rotate
the handpiece until contacts align, then fully insert
handpiece.
•
Hold empty handpiece in a semi-upright position over a
sink or drain. Activate the Tap-On
water exits to release any air bubbles that might
be trapped inside the handpiece. NOTE: Tap-On
Technology only operates when an insert is in the
handpiece.
•
Lubricate the O-ring on the insert with water before
placing it into the handpiece. Fully seat insert with a
gentle push-twist motion. DO NOT FORCE.
•
Turn the Lavage Control to select flow rate during
system operation. Flow rate is placed on a scale from 1
to 6. Turn control clockwise toward 6 to increase flow
at insert tip. Turn control counter-clockwise toward 1
to decrease flow. The flow rate through the handpiece
also determines the temperature of the lavage. Lower
water flow rates produce warmer lavage. Higher flow
rates produce cooler lavage. If the handpiece becomes
14
Foot Pedal until
TM
TM