Incoming water temperature to the Cavitron System
should not exceed 25˚C (77˚F). If needed a device
should be installed to maintain a temperature within this
specification, or a Cavitron DualSelect Dispensing
System attached to allow this system to be operated
as a closed water system.

7 2 Electrical Requirements

Incoming power to the system must be 100 volts AC to
240 volts AC, single phase 50/60 Hz capable of
supplying 1.0 amps.
The system power should be supplied through the AC
power cord provided with your system.
WARNING: To avoid risk of electric shock, this equipment
must only be connected to a supply mains with protective
earth.

7 3 Unpacking the System

Carefully unpack your Cavitron Plus Ultrasonic Scaler and
verify that all components and accessories are included:
1. Cavitron
Plus
System with Handpiece Cable
®
Assembly with swivel
2. Water Line Assembly (Blue) with Filter and Quick
Disconnect
3. Additional Water Line Filter
4. Detachable AC Power Cord
5. Cavitron
Tap-On
®
TM
6. "AA" Batteries (4-Pack)
7. Auxiliary Cable for Tap-On
8. Steri-Mate
Grip (not shown)
®
9. Steri-Mate
Detachable Sterilizable Handpiece
®
10. Cavitron
Ultrasonic Inserts (quantity optional)
®
11. Efficiency Indicator for Cavitron Inserts
12. Literature Packet

7 4 System Installation

The Cavitron Plus System is designed to rest on a level
surface. Be sure unit is stable and resting on four feet.
The Cavitron Plus System should not be positioned such
that access to the power input and AC power cord are
limited.
Wireless Foot Pedal
Foot Pedal
TM
Placing unit in direct sunlight may discolor plastic
housing.
The system has been equipped with a Tap-On
Foot Pedal which was factory synchronized to operate
with the system's base unit. If your office has more
than one Cavitron system with Tap-On
it is recommended that you mark the Tap-On
Pedal and base unit for easy reference as to which Tap-
On
Foot Pedal operates with which base unit. Should
TM
resynchronization be necessary, follow the instructions in
section 7.10.

7 5 Power Cord Connection

Verify the Main Power ON/OFF switch, located at the
center front underside of the System, is set to the OFF
(O) position before proceeding.
Insert the AC power cord into the power input on the
back of the System.
Insert the pronged plug into an AC wall outlet.

7 6 Water Supply Line Connection

Grasp the Water Supply Line (blue hose) by the end
opposite the quick-disconnect and insert it into the water
inlet connector until fully seated.
9
Wireless
TM
Technology,
TM
Foot
TM
Press ring to release
water supply tube.