8.6
Starting the sterilization cycle
Select the sterilization cycle most suitable to the loaded materials by pressing the button
B-SELECT
(FIG.2) Each single pressure changes the program selection to the next one.
Programs are displayed After the program selection, start the cycle by pressing the button
Start
During the cycle the display shows all parameters and information related to the cycle in
progress. In this setup the display shows: the type of cycle, the cycle status, the time
remaining before the end of cycle (for the Vacuum Test it identifies the whole cycle, while
for all the other cycles it identifies the sterilization phase plus that of drying), the number of
cycles done by the machine and the button
parameters (FIG.26).
8.7
End of cycle
An acoustic signal and a displayed image warn that the sterilization cycle has been
completed. Press any button to continue and unlock the door by pressing one of the
buttons Unlock. If there is pressure inside the chamber the button will not activate the
release device. When the door is unlocked pull the door handle and open.
8.8
Unloading the sterilized materials
Wear personal protective equipment in compliance with regulations on safety and hygiene
at work. Extract the trays by using the spanner provided (FIG.10), let rest the instruments
and store them in environments which are not exposed to contamination.
26
Info
allowing access to the list of working
FIG.26