Night Cycle - Tecno-gaz ANDROMEDA PLUS xp Manual De Uso

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

PAR. 12

NIGHT CYCLE

The autoclave is fitted with a special economizer interrupting heating after 60 inactivity minutes (STAND BY).
This device makes it possible to carry out sterilisation cycles even in the absence of the operator, because
once the cycle has been ended after 60 minutes, the autoclave is automatically switched off.
All the LEDS are turned off and only the main ON/OFF switch remains on.
To check that sterilisation has taken place, the operator must press button "0/1".
In case the "TIME/MESS." display indicates "END", this means that the sterilisation has taken place; if an
alarm or error message is displayed, see PAR. 15.
In short:
In case by pressing the button "0/1", on the "TIME/MESS." display, an alarm code is visualised, it is
necessary to repeat the sterilisation cycle, because during the execution an error has taken place.
In case the autoclave remains switched on with opened door or without having set the sterilisation cycle, the
autoclave is automatically set on the "STAND BY" phase for 60 minutes.
Afterwards, by resetting the functions, pressing keys 11A, on the "TIME/MESS." display, the message of the
previously set cycle will be visualised.
In case a failure takes place during the sterilisation cycle, it is immediately signalled on the "TIME/MESS."
display.
To reset the functions, it is necessary to press "0/1" , to select once again the wished cycle and to start by
pressing button "0/1".