8.3

Clean water tank filling

Connect the water filling hose with filter (FIG.15), and fit it into the front coupler of the
autoclave (FIG.2- pos.9).
Fit the other end of the hose with filter inside the tank of demineralized or distilled water.
Now press the button
is not achieved within 180 seconds the pump stops automatically and it will be necessary
to press button
B-PUMP
when the maximum level is achieved.
If the autoclave is mounted with the demineralization system comply with the instructions
shown in the package of this device and those of chapter 16.
8.4

Characteristics of the water to be used

TABLE SHOWING THE QUALITY LEVELS LAID BY THE UNI EN 13060 : 2009
Evaporation residue
Silicon oxide
Iron
Cadmium
Lead
mg/l Remains of heavy metals apart from iron,
cardium, lead
Chloride (Cl')
Phosphate (P20s)
Conductivity (at 20° C)
Ph value (acidity level)
Appearance
Hardness (and alkaline earth ions)
24
B-PUMP
(FIG.2) to start the water filling pump. If the maximum level
again to complete the tank filling. The pump stops automatically
STANDARD UNI EN 13060 : 2009
10
1
0.2
0.005
0.05
0.1
2
0.5
15
5 ÷ ÷ ÷ ÷ 7,5
Transparent, clear, without
deposits
0.02
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
µs/cm
mmol/l