Messages of error are emphasized through an Alfa-numerical code, consisting in a letter
and three numbers.
The "A" code is relative to alarms, irregularities of the autoclave, after having carried out
the remedy, if the problem persists it is necessary to request telephonic assistance.
If an alarm message is displayed (CODE "A") the cycle is to be considered
aborted: it will be necessary to repeat all the preparation and sterilization
operations.
To reset alarms and errors hold down at the same time the buttons under the bar
indicated with Reset.
ERROR
A 001
Cycle interrupted by the user
A 101
Vacuum not achieved in 10 min.
A 111
Vacuum not maintained on the first phase of VACUUM TEST
A 121
Vacuum not maintained on the second phase of VACUUM TEST
During the warm-up phases the tool has not charged the correct amount
A131
of water
A132
Error in the flow-meter operation
Pressure above the allowable limit during the water recall for the warm-
A133
up phases
A 200
Error in the control of the SV operation
A 400
Error in the locking operation
A 401
Error in the locking operation
A 403
Error in the locking operation
A 405
Error in the locking operation
A 551
Pressure out of limit
A 637
Error in the access to the memory card
A 651
Reading of the T1 probe during sterilization above the maximum limit
A 653
Reading of the T2 probe during sterilization above the maximum limit
A 661
Error in the probe reading
A 662
Error in the probe reading
Error for failed pressure achievement during the first two phases of the
A 701
warm-up
MESSAGES OF ERRORS
CAUSE
ENGLISH
35