7. STERILISATION TABLE
CYCLES
(not wrapped)
(not wrapped)
The choice of the sterilisation cycle must be based on the data supplied by the producer of the
object you want to sterilise.
8. FILLING OF THE WATER TANK
Take the cover away from the water tank. The cover is the punched plate with the knob.
Fill in the tank with distilled or demineralised water. Pour water until the lower edge of the plate
divider Photo B.
Attention: the water in the tank must not top the plate divider.
Put the cover on the water tank.
NOTE: the use of feedwater with contaminants at levels exceeding those given in the table
above can greatly shorten the working life of the device, with damage at the components of
the autoclave and then invalidate the manufacturer's warranty!
STERILISATION
TIME
[minutes]
121° C
22,5
134° C
8
DRYING
TIME
[minutes]
5
5
Photo B
WORKING
PRESSURE
[Bar Relative]
1,3 ± 0,2
2,25 ± 0,2