6.2.2
Option 2: morcellation with Trocar Cannula
1.
Hysterectomy
Select the procedure for the hysterectomy in ac-
cordance with current medical standards.
2.
Inserting Trocar Cannula
Insert Trocar Cannula with Obturator into an ex-
isting access under visual control (endoscopy).
Remove Obturator.
3.
Insert the cutting tube through the Trocar Cannula.
Actuate the valve flap lever of the Trocar Cannula
and hold in this position.
Carefully insert prepared cutting tube with gear
unit and seal unit into the proximal opening of
the Trocar Cannula.
Cutting tube must be fully inserted.
Release the valve flap lever of the Trocar Cannula.
The guide pin of the Trocar Cannula must be in
the distal slot of the gear unit.
Lock the Trocar Cannula in position with a click.
Make sure that the cutting tube remains in the
NO CUT position.
4. Inserting gripper
Insert gripper carefully through the proximal
opening of the gear unit and Trocar Cannula.
5.
Preparing cutting tube
Rotate the Trocar Cannula so the CUT position is
visible from above.
If the cutting tube is completely inserted, the cut-
ting edge is now only partially covered by the
Trocar Cannula.
24
Instrucions for use for TCM 3000 BL - Morcellator, Ref. 31946, V26/16
TCM 3000 BL - Morcellator
CUT