10. Connecting motor
Insert the motor coupling into the bottom end of the
handle.
Continue to insert the motor coupling until it locks into
the gear unit motor connection.
Align the red dot on the device connection plug of the
motor with the red dot on the motor connection of the
control unit.
Plug the motor device connection plug into the motor
connection of the control unit. The plug must lock in po-
sition.
11. Connecting pedal
Align the red dot on the device connection plug of the
pedal with the red dot on the pedal connection of the
control unit.
Plug the pedal device connection plug into the pedal
connection of the control unit. The plug must lock in
position.
Functional check
Check Morcellator for loose parts
Check that Morcellator is correctly assembled
Are all labels easily readable?
Check cutting tube: cutting tube evenly ground, no broken edges or bends?
Switch on control unit
Switch on motor (press pedal).
Run motor with Morcellator for about 20 seconds; it must reach the set speed.
If it does not run evenly, switch off motor with Morcellator.
Switch off control unit.
The control unit will only operate if the pedal has been connected to the control unit.
Check that Morcellator is correctly assembled.
Check that the gear unit runs smoothly; if necessary treat it with Nou-Clean care spray.
Check that the motor is clean and runs smoothly; if necessary clean it and treat with Nou-Clean care
spray.
Repeat the function check.