TCM 3000 BL - Morcellator
6.5
Disassembly of motor
Disassembling motor (three steps)
Disassemble motor before decontamination.
1.
Unscrew handpiece holder from motor
2.
Unscrew motor cap from motor
3.
Pull motor plug from motor
6.6
Disassembly of gear unit and Morcellator with protective tube
Disassembling seal unit
Remove membrane screw from seal unit and
remove membrane seal.
Remove seal retainer from gear unit and remove
roof shaped seal.
Risk of injury for the user!
The distal end of the cutting tube is very sharp, handle the cutting tube very carefully.
Removing protective tube and cutting tube
Make sure that the lock nut of the protective
tube covers the distal end of the cutting tube
(protected NO CUT position is visible).
Hold the gear unit with one hand so the distal
end of the cutting tube does not point to the user
or any other person.
With the other hand carefully disconnect the pro-
tective tube from the gear unit and pull it off.
Hold the cutting tube with the same hand and
press it lightly against the gear unit until it re-
leases itself from the holder and slides out
smoothly. It can now be pulled out of the proxi-
mal opening of the gear unit.
Instrucions for use for TCM 3000 BL - Morcellator, Ref. 31946, V26/16
1. Remove pro-
2. Pull out cut-
tective tube
ting tube
EN
29