7.4

Preparation of gear unit

The gear unit is designed for 250 sterilization cycles.
Oil may leak from the pressure compensation opening of the gear unit under pressure.
Wipe off with a moist cloth.
Do not immerse gear unit in liquids.
Disassembly
Disassemble gear unit as described in section 6.6 "Disassembly of gear unit and Morcellator".
Manual cleaning
Use neutral and alkaline cleaning agents, cleaning brushes and deionized water.
Rinse and brush surfaces and intermediate spaces thoroughly with neutral and alkaline cleaning agent.
Brush and rinse lumen thoroughly with neutral and alkaline cleaning agent.
Rinse all surfaces and lumina with deionized water for at least 10 minutes. Make sure that all intermedi-
ate spaces are cleaned.
Machine cleaning and disinfection
Equipment: Cleaner and disinfector with special racks that ensure that channels are cleaned. Use only
neutral cleaning agents.
Place the gear unit in the load carrier (ensure that the channels can be rinsed).
Place seal holder and membrane screw in basket.
Set a cleaning cycle that offers sufficient cleaning and rinsing. Perform the final rinse with deionized
water.
Perform a 10-minute rinse cycle at 93°C to facilitate thermal disinfection.
When removing the gear unit, the seal holder and the membrane screw, check whether visible contami-
nation still remains in the gaps and the grooves. If necessary, repeat the cycle or clean manually.
Maintenance, lubricating gear unit
Perform a visual inspection to check for dam-
age, corrosion and wear and tear.
Use Nou-Clean spray exclusively.
Attach the spray attachment for handpieces
(Ref. 1928) to Nou-Clean spray and spray gear
unit for about three seconds and briefly wipe
with a moist cloth (see instructions on spray
can).