DENTSPLY DeguDent Cercon heat plus P8 Instrucciones De Uso página 40

Idiomas disponibles
  • ES

Idiomas disponibles

6 Operation
40
To start a program, the following conditions must be met:
• The temperature inside the sintering chamber is less than
200 °C.
• The display indicates that the lifting table was ready to be
lifted by ("rel1") and has been lifted all the way.
• No program LED (green) is lit.
• No error message is shown (see Error messages and
troubleshooting).
• Lifi tng table is lifted up
Note: If the sintering chamber temperature exceeds
200 °C, the lifting table can no longer be lowered.
6.3 Canceling program execution
Program execution can only be canceled while the tem-
perature is below 200°C.
To do so, press the STOP key. The heating process is
discontinued, and the green program key LED turns off.
6.4 Program execution
While the program is running, the calculated remaining ex-
ecution time is shown on the display (hh:mm format).
The display shows "rel2" to indicate that the heater has
been turned on; during the heating process, the display
also shows "heat."
While the program is running, the DISPLAY key can be
used to alternate between execution time and temperature
display (see DISPLAY key, INFO key).
loading