en
6.10 Steam sterilization
Damage or destruction of the
batteries due to processing!
CAUTION
Note
Remove all attached components from the product
(tools, accessories) before sterilization.
Check to ensure that the sterilizing agent will come
into contact with all external and internal surfaces
(e.g. by opening any valves and faucets).
Use a validated sterilization method:
– Steam sterilization using fractional vacuum
process
– Steam sterilizer DIN EN 285 and validated pur-
suant to DIN EN ISO 17665
– Sterilization using fractionated vacuum process
at 134 °C/holding time 5 min
When sterilizing several products at the same time in
one steam sterilizer:
Ensure that the maximum permitted load specified
by the manufacturer for the steam sterilizer is not
exceeded.
6.11 Storage
Store sterile products in germ-proof packaging,
protected from dust, in a dry, dark, temperature-
controlled area.
18
Do not sterilize the
rechargeable battery.
7.
Maintenance
To ensure reliable operation, the product must be
maintained once per year as indicated on the mainte-
nance label, see Fig. 11, e.g. February 2012.
Fig. 11
For technical service, please contact your national
B. Braun/Aesculap agency, see Technical Service.