en
6.10 Steam sterilization
Damage or destruction of the
batteries due to processing!
►
CAUTION
Note
Remove all attached components from the product
(tools, accessories) before sterilization.
►
Check to ensure that the sterilizing agent will come
into contact with all external and internal surfaces
(e.g. by opening any valves and faucets).
►
Use a validated sterilization method:
– Steam sterilization using fractional vacuum
process
– Steam sterilizer DIN EN 285 and validated pur-
suant to DIN EN ISO 17665
– Sterilization using fractionated vacuum process
at 134 °C/holding time 5 min
When sterilizing several products at the same time in
one steam sterilizer:
►
Ensure that the maximum permitted load specified
by the manufacturer for the steam sterilizer is not
exceeded.
6.11 Storage
►
Store sterile products in germ-proof packaging,
protected from dust, in a dry, dark, temperature-
controlled area.
18
Do not sterilize the
rechargeable battery.
7.
Maintenance
To ensure reliable operation, the product must be
maintained once per year as indicated on the mainte-
nance label, see Fig. 11, e.g. February 2012.
Fig. 11
For technical service, please contact your national
B. Braun/Aesculap agency, see Technical Service.