en
6.3
Preparation at the place of use
►
Remove all attached components from the product
(tool an accessories).
►
Remove any visible surgical residues as much as
possible with a damp, lint-free cloth.
►
Place the dry product in a sealed waste container
and forward it on for cleaning and disinfection
within 6 hours.
6.4
Preparation before cleaning
►
Prior to the first mechanical cleaning/disinfection
process, mount the ECCOS fixations in a suitable
tray (e.g. GB243800).
►
Insert the product in its proper position in the
ECCOS fixation.
14
6.5
Cleaning/disinfection
Product-specific safety instructions for the
reprocessing procedure
Damage to the product due to
inappropriate cleaning/disin-
fecting agents and/or exces-
CAUTION
sive temperatures!
►
►
►
►
►
►
Damage or destruction of the
batteries due to processing!
►
CAUTION
Note
The indicated drying time is a guide time only. It must be
checked taking into account the specific conditions (e.g.
load) and if applicable adjusted.
Use cleaning and disinfect-
ing agents according to the
manufacturer's instruc-
tions which
be approved for plastic
material and high-
grade steel,
do not attack softeners
(e.g. in silicone).
Do not use cleaning agents
that contain acetone.
Observe specifications
regarding concentration,
temperature and exposure
time.
Do not exceed the maxi-
mum temperature of 60 °C
during chemical cleaning
and/or disinfection.
Do not exceed the maxi-
mum temperature of 96 °C
during thermal disinfection
with FD water.
Dry the product for at least
10 minutes at a maximum
of 120 °C.
Protect battery from mois-
ture.