Adjusting the sheath

The silicone sheath should be 5 - 11 cm shorter than the optical fiber itself. If
the sheath is too short, the optical fiber may not be fastened properly in the
handpiece. If it is too long, the optical fiber cannot be inserted in the hand-
piece. To shorten the sheath to the correct length, proceed as follows:
Hold the optical fiber approx. 11 cm below the distal end.
Press the sheath and the optical fiber together.
Strip approximately 10 cm of the sheath with the fiber cutter.
i
NOTICE
After stripping the optical fiber, bend it back and forth to make sure it has not
been damaged.
4.8.5

Adjusting the fiber with the fiber cutter

Removing the sheath
1.
Place the fiber in the guide suitable for its thickness.
2.
Press the cutter together at its open end.
3.
Keep pressing the fiber cutter and pull off the sheath.