CAUTION
If you cannot see the red aiming beam after switching the laser on or during
treatment, proceed as follows:
Check to see if the optical fiber or its connector is damaged.
If the optical fiber is damaged, replace it with a new one.
Connect the optical fiber to the laser unit.
Use the optical fiber to check whether the aiming beam signal is visible.
If you cannot detect any damage to the optical fiber and the signal of the
aiming laser is not visible with the new optical fiber either, switch off the
laser and contact Sirona Dental Systems GmbH, your local dental depot
or your authorized service center.
WARNING
Use of the laser unit when the aiming beam is not functioning properly may
cause injuries to operating personnel, assistents or patients.
CAUTION
The optical fiber must be cleaned and disinfected prior to the first treatment.
CAUTION
Remove the protective caps only for operation.
CAUTION
Never touch the connector ends and protect them against damage.
CAUTION
Never bend, fold or jam the optical fiber, as this might cause it to break.
CAUTION
Never pull on the optical fiber.
CAUTION
Bear in mind the maximum bending radius of the optical fiber:
• Short-term (during treatment): 100 x radius of optical fiber
• Long-term (during storage): 600 x radius of optical fiber
i
NOTICE
The unit detects whether the optical fiber is connected via the magnet.