60 56 092 D 3485
D 3485.201.01.10.02 03.2008
Laser irradiation (and hence treatment) can be terminated immediately by
pressing the red "LASER STOP" emergency button.
To quit the selected preset menu, press the "ENTER" key.
A message stating that therapy has been terminated appears on the
display.
The "LEFT ARROW" key returns you to the main menu.
5.8
"MANUAL SETTING AREA" submenu
After the "MANUAL SETTING AREA" is called up, a figure with configuration
settings appears on the display (see Fig. 24).
Here you can select from ten different programs with the arrow keys.
Pressing the "ENTER" key takes you to the selected program
(PP1 for program 1, PP2 for program 2, etc.)
Confirm the optical fiber you would like to use for your personal settings
(200µm or 320 µm) with "ENTER", or change this selection with the
"RIGHT ARROW" key.
The following happens after you select the submenu:
The green LEDs start flashing, indicating that the laser is in standby mode
now.
The laser is now ready for operation.
Wear protective goggles whenever the aiming beam is activated.
Figure 24:
"MANUAL SETTING AREA" submenu
Selected program: in this example P3:
1.
2.
Laser activation time in milliseconds: in single pulse or auto-repeat
mode. In the autorepeat mode, the laser unit switches back on after
two seconds if the foot switch is actuated. If set to any time, laser radiation
will stop after this time, even if you continue to press the foot control or finger
switch. If set to zero, "------" appears on the display, and the laser remains
activated as long you press the foot control or finger switch. You can set a
time between 100 ms and 60 s (60,000 ms) in 50 ms intervals. When you
press the key for more than 3 seconds, the value increases at a faster rate.
5 Operating Instructions
37