Interlock; Explanation; Installation Of An Interlock With Door Switch - Sirona Dental SIROLaser Instrucciones De Uso

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

INTERLOCK/SWITCH
Figure 2:
Connection of the Y
connector with the laser
system
Figure 3:
Connection of the interlock
plug with the Y connector
60 56 092 D 3485
D 3485.201.01.10.02 03.2008
4.4

Interlock

4.4.1

Explanation

The interlock is a safety device that stops laser radiation whenever the door
of the treatment room is opened. The interlock circuit must be connected to a
switch that is located near the door of the treatment room in order to ensure
automatic interruption of the laser emission.
i
NOTICE
The installation must be performed by a qualified electrician who is also re-
sponsible for the installation and maintenance of the electrical system to which
the SIROLaser is connected.
Please request our technical data sheet with circuit diagram for the installa-
tion of the interlock circuit (Service manual on safety checks, REF 6196542).
i
NOTICE
Additional or different safety precautions required by the applicable national or
local regulations for the protection of dentists, assistant personnel, or patients
must also be observed.
4.4.2

Installation of an interlock with door switch

Plug the Y connector into the "INTERLOCK/SWITCH" socket.
As soon as the cable is connected to the interlock connector, connect the
interlock connector to the Y connector.
4 Installation
19