3 Introduction
14
WARNING
The laser vapor contains tissue particles. Always wear a face mask, as a risk
of infection otherwise exists.
Please also observe the following:
i
NOTICE
The unit may be operated only in rooms that comply with the requirements of
Council Directive 89/336/EEC. Any applicable national or local legal require-
ments must be complied with.
CAUTION
Avoid interference between the laser emission and any optical sensors of de-
vices operated in the vicinity of the SIROLASER.
3.3
Transport and storage
3.3.1
Transport and storage
The SIROLaser comes in a case that ensures proper and easy transport.
Please be sure to observe the following:
CAUTION
Do not leave the laser unit in a vehicle parked in the sun. The inside tempera-
ture of the car could thus heat up to a point where individual components may
be damaged.
To ensure appropriate storage, the device must always be kept in the case
supplied by Sirona Dental Systems.
Thus stored, the SIROLaser can withstand the following ambient conditions:
Temperatures from -40°C to +70°C
Relative humidity from 10% to 90%
Atmospheric pressure from 800 hPa to 1060 hPa
In its original transport packaging, the SIROLaser can withstand the following
ambient conditions:
Temperatures from -40°C to +70°C
Relative humidity from 10% to 95%
Atmospheric pressure from 800 hPa to 1060 hPa
3.3.2
Operating conditions
The SIROLaser may be operated in the following environmental conditions:
Temperatures from +10 °C to +33 °C
Relative humidity from 10% to 95%
Atmospheric pressure from 800 hPa to 1060 hPa
CAUTION
Following transport and storage, let the laser unit adapt to room temperature
for about one hour prior to operation to reduce the risk of malfunctions caused
by condensation.
60 56 092 D 3485
D 3485.201.01.10.02 03.2008