4. Technical data related to electromagnetic compatibility (EMC)
EOlife® is designed to operate in an electromagnetic environment as defined below. The user should ensure that the
device is used in such an environment
Interference measurements
EN
Conducted emissions – Mains
terminal disturbance voltage
according to CISPR11
Radiated emissions – Radiation
disturbance according to
CISPR11
Harmonic current emissions
according to IEC 61000-3-2
Voltage fluctuations/flicker
according to IEC 61000-3-3
EOlife® is designed to operate in an electromagnetic environment as defined below. The user should ensure that the
device is used in such an environment
Immunity tests
Electrostatic discharge (ESD)
according to IEC 61000-4-2
Electrical fast transients/bursts
according to IEC 61000-4-4
Shock waves/surges according
to IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations according to IEC
61000-4-11
Conducted disturbances
according to IEC 61000-4-6
Power frequency magnetic
field according to IEC 61000-
4-8
Radiated disturbances
according to IEC 61000-4-3
EOlife® has been tested for immunity to the radio services listed below. If the field strength measured on the site where
EOlife® is used exceeds the level of compliance below, it is advisable to monitor EOlife® to ensure that it is functioning
properly If this is not the case, it may be necessary to take additional measures, such as changing the direction or
location of EOlife®. Specific examples of RF transmitters known to cause electromagnetic interference are diathermy
and electrocautery devices, RFID devices and security systems (such as electromagnetic antitheft devices and metal
detectors) Some of these RF transmitters (such as RFID devices) may be concealed in the area where the device is to be
used, potentially exposing it to these RF transmitter fields.
48
Manufacturer statement and guidance – Electromagnetic interference
Compliance
Class B: 150 kHz to 30 MHz
Class B: 30 MHz to 1 GHz
Class A 0 kHz – 2 kHz
0 kHz to 2 kHz
Manufacturer statement and guidance – Electromagnetic immunity
IEC 60601 test level
Contact: ±8kV
Air: ±2kV ±4kV, ±8kV, ±15kV
Access by power
AC/DC:
± 2 kV, f = 100 kHz
Access by AC/DC power ports:
± 0 5 kV, ±1 kV (Line/Line)
Min & max voltage:
0% UT for 0.5 cycle
0% UT for 1 cycle
70% UT for 25/30 cycles
0 % UT for 250/300 cycles
Access by AC/DC power ports:
3 V (0 15 MHz to 80 MHz)
6 V (+ ISM bands)
80% MA, 1 kHz
30 A/m
50 Hz or 60 Hz
80 MHz to 2 7 GHz
AM modulation 80%, f = 1 kHz
10 V/m: Home
Electromagnetic environment – Guidance
Equipment suitable for use in households
and buildings used for domestic purposes
directly connected to a low voltage power
supply
/
/
Level of compliance
Contact: ±8 kV
Air: ±15 kV
5/50 ns, 100 kHz; ±2 kV
1.2/50 (8/20) μs LtL: ±1.0 kV
0% UT for 0.5 cycle (1 phase)
0% UT for 1 cycle
70% UT for 25/30 cycles
0% UT for 250/300 cycles (50/60 Hz)
150 kHz – 80 MHz 3 V
ISM/amateur bands 6 V
80% / 1 kHz
other as defined in RMF
30 A/m 45
50 Hz or 60 Hz
80 MHz to 2 7 GHz
Home Healthcare 10 V/m Prof Healthcare
3 V/m
80% / 1 kHz
other as defined in RMF