X. Technical data
1. Technical data relating to the device
Specifications
Product class according to
Regulation (EU) 2017/745
Dimensions (W x H x D)
Weight including battery
Operating conditions
Transient operating conditions
(maximum 20 min)
Storage/transport conditions
Lifetime
Power supply (nominal
voltage)
Run time
Classification according to EN
60601-1:
o Type of protection against
electric shock
o IP rating for protection against
electric shock
IP rating for protection against solids,
dust and ingress of water
Electromagnetic compatibility
(EMC) according to EN 60601-1-2
BLE module frequency range
Signal strength
Certification
Shock and vibration resistance
Screen
135 mm x 75 mm x 35  mm
150g ± 5  g
o Temperature between 0°C and +40°C
o Relative humidity between 15% and 95% (non-condensing)
o Atmospheric pressure range of 620 hPa (at an altitude of 4,000 m) to 1,060
hPa (at an altitude of -500 m)
o Temperature between -20°C and +50°C
o Relative humidity between 15% and 90% (non-condensing)
o Temperature between -40°C and +70°C
o Relative humidity between 10% and 90% (non-condensing)
5 years
3 7 V
5 hours
The entire device except for the charger (electronic unit, battery and
FlowSense® sensor) has been designed to meet the requirements for type BF
applied parts
IP44 (in use, i e when EOlife® is connected to its battery and to the
FlowSense®)
The control parameters and threshold values can be obtained from the
manufacturer
2,402 – 2,480 GHz
Max 4 dBm
Directive 2014/53/EU, FCC, IC, MIC, KC
EN 60601-1-12 (category: resistance in an emergency vehicle)
60 x 45 mm
Resolution 320 x 240 pixels
EOlife®
Class I
EN
45