Switch off EOlife® by pressing the ON/OFF button, then on
8
the OFF icon or by pressing the ON/OFF button for 3 seconds
EN
Switching off using
the summary screen
IX. Detailed product description and operating instructions
EOlife® is controlled via the touch screen and the physical ON/OFF button
Caution
Avoid using any object that could scratch the screen
1. Using the device
a) Assembly/disassembly
Assembly
Remove the FlowSense® sensor from its packaging, using the notch to tear the antistatic bag to avoid damaging the
sensor
Caution
Check the sensor's expiry date before using. The use by date can be found on the label affixed to the bag ( ) Check
that the bag is intact before using the FlowSense® sensor If the FlowSense® sensor bag is torn, discard the sensor
and use another one as water or contaminants may have entered the bag
Connect the FlowSense® sensor to the EOlife® unit, taking care not to expose the printed circuit board to dust or
splash it with water
First insert the part of the FlowSense® with a notch into the connector ridge, and then press firmly on the sensor
so that it comes into contact with the connector pins You will hear a click when the connector tab holds the
FlowSense® sensor in place
Check that the FlowSense® sensor is held properly in place before starting up the EOlife® unit
Tab
34
Notch
Ridge
Disconnect the ventilation bag and interface
9
(mask or endotracheal tube) and unclip
the FlowSense® sensor by pulling the tab
outwards, as shown by the arrow
Disconnecting EOlife®
click