Alarm message linked to the FlowSense® sensor
During ventilation, if the FlowSense® sensor is defective or not properly connected or if a foreign object on the
connector interferes with the signal transmission, an alarm screen is displayed:
You must reconnect a functional FlowSense® sensor as described above to clear the "FlowSense failure" alarm.
Note
The "FlowSense failure" alarm screen can only be displayed from the main screen.
EOlife® device failure
If there is an EOlife® device failure, the red LED in the top left corner of the screen comes on This red LED tells you
when there is a device failure whether it is on or off
Warning
Do not use the EOlife® device if the red LED stays on
See "XI. What to do in the event of an incident"
d) Standby mode
If there is no user activity (no ventilation or no action on EOlife®) for 3 minutes, the device goes into «standby» mode
and the screen turns off
To turn the screen on again, simply press the physical ON/OFF button and the summary screen will reappear
Note
All settings are maintained after the standby mode
If EOlife® remains in standby mode for more than 10 minutes, the device will turn itself off completely
2. EOlife® device set-up screen
a) Set-up screen display
1.
Start up EOlife® by pressing the physical ON/OFF button
2. Press normally the physical ON/OFF button to show the summary screen
FlowSense failure
Alarm screen
"FlowSense failure"
Summary screen
41