Eliminating faults or malfunctions

Eliminating faults or malfunctions

Fault table

Faults are shown on the torque wrench display with an alternating fault code and a warning triangle. Faults
are additionally indicated using optical and acoustic signals.
Fault code
E1
Not adjusted LR
No serial number
E3 and ">0<"
The torque wrench
is switched off
again after
switching on.
E4
E5
22
Signals
Cause
Red LED flashing
Electronic fault
Electronic fault
Electronic fault
Red LED flashing
The torque wrench is loaded
during taring.
The measuring element is
damaged
Red LED flashing
The battery is empty.
Yellow LED
The battery is almost empty.
flashing
Red LED flashing
The torque wrench has been
overloaded.
Red LED flashing
Calibration is due.
Remedy
Send the torque wrench to
the manufacturer for repair.
Send the torque wrench to
the manufacturer for repair.
Send the torque wrench to
the manufacturer for repair.
Switch the torque wrench off
and on again.
Tare the torque wrench
manually.
If the fault message persists,
send the torque wrench to
the manufacturer for repair.
Replace the battery with a full
battery of the same type.
Replace the battery with a full
battery of the same type in
the near future.
Calibrate the torque wrench
or have it calibrated.
Use above 110 % is to be
avoided.
Calibrate the torque wrench
or have it calibrated.