Section 6: Infection Control

6.1 General Information

• For operator and patient safety, carefully practice the
infection control procedures detailed in the Infection Control
Information Booklet accompanying your System. Additional
booklets can be obtained by calling Customer Service at
1-800-989-8826, Monday through Friday, 8:00 A.M. to 5:00 P.M.
(Eastern Time). For areas outside the U.S., contact your local
DENTSPLY Professional representative.
• As with high speed handpieces and other dental devices, the
combination of water and ultrasonic vibration from the Cavitron
Built-In Scaler will create aerosols. Following the procedural
guidelines in Section 10 of this manual can effectively control and
minimize aerosol dispersion.

6.2 Water Supply Recommendations

• It is highly recommended that all dental water supply systems
conform to applicable CDC (Centers for Disease Control and
Prevention) and ADA (American Dental Association) standards,
and that all recommendations be followed in terms of flushing, and
general infection control procedures. (See Sections 7.2 and 11)
Knowledge of and compliance with agency guidelines, standards and
recommendations is the sole responsibility of the Dental Healthcare
Professional.
• As a medical device, Cavitron
®
products need to be installed
in accordance with local or national regulations, including guidelines
for water quality (e.g. drinking water). As an open water system,
such regulation may require your Cavitron product to be connected
to a centralized water control device that prevents water containing
contaminates from back-flow into the water supply.
ENGLISH • 6
All manuals and user guides at all-guides.com
Section 7: Installation Instructions

7.1 Installation

Your DENTSPLY Cavitron
your dental treatment unit. The power source is double insulated type SELV device and qualified to IEC 601.1.
®
Built-In Scaler has been installed by trained technicians. The diagram below shows all connections to
UNIVERSAL CONTROL BOARD
J3M-1
NO CONNECTION
J3M-2
NO CONNECTION
J3M-3
NO CONNECTION
J3M-4
POWER CONTROL- HIGH
J3M-5
POWER CONTROL- WIPER
J3M-6
NO CONNECTION
J3M-7
ULTRASONIC ON/OFF
J3M-8
POWER CONTROL- COMMON
POWER DRIVE BOARD
J1-1
HANDPIECE- DRIVE
J1-2
HANDPIECE- GROUND
J1-3
HANDPIECE- FEEDBACK
J2-1
24 VAC
J2-2
24 VAC
J3-1
POWER RESISTOR
J3-2
POWER RESISTOR
VIOLET
YELLOW
BROWN
BLACK
RED
GREEN
WHITE
BLUE
BLUE/WHITE STRIPE
BLACK
BLACK
ENGLISH • 7