Severity
Alarm Header
Complete Procedure
Controller Error
Impella
Supported
Impella Defective
Placement Signal Not
Reliable
Purge Cassette Failure
Purge Volume Critically
Low
Reinstall Software
Suction
AC Power Disconnected
Audio Off
Preventing Retrograde
Flow
Purge Cassette
Incompatible
24
Table 4 Automated Impella
Action
1. Follow the steps on the screen or
2. Exit the procedure
Switch to backup controller.
Catheter Not
1. Replace Impella with supported
®
catheter (2.5, CP, 5.0, LD, RP).
2. Contact Abiomed Service to upgrade
Impella Controller.
Do not use Impella.
Replace Impella.
Monitor patient hemodynamics
Replace purge cassette.
1. Open the PURGE SYSTEM menu
and select Change Purge Fluid.
2. Follow the instructions to change
the purge fluid.
Software installation was unsuccessful.
Reinstall software.
1. Check filling and volume status.
2. Check Impella position.
3. Reduce P-Level.
Controller is running on battery power.
The audio for the following alarms has
been disabled.
<Alarms will be listed here>
Impella P-level has increased to prevent
retrograde flow.
1. Consider increasing target P-level.
2. For weaning, disable Retrograde Flow
Control through MENU soft key.
Contact Abiomed Service to update
Impella Controller.
®
Controller Alarm Messages (continued)
Cause
User has not responded to a de-air
or purge procedure screen for more
than 1 minute or a transfer to
standard configuration screen for
more than 5 minutes.
There is a problem with the controller
electronics.
The Impella
Catheter is not
®
supported to operate with the
current version of Automated
Impella
Controller software and/or
®
hardware.
There is a problem with the Impella
RP
Catheter electronics.
®
There is a problem with the Impella
Catheter sensor signal.
There is a problem with the purge
cassette hardware.
There are 15 mL (in addition to 5%
of the starting bag volume) or fewer
remaining in the purge fluid bag.
Software was not installed
successfully.
Suction is detected.
AC power was disconnected.
User has disabled audio for
Placement Signal Not Reliable,
Suction, Purge Pressure High or
Purge System Blocked alarm.
Retrograde flow has been detected
and minimum motor speed has been
increased to more than target P-level.
Incompatible purge cassette RFID
version.