10. If you have not already done so, disconnect and discard the Y connector with the red
and yellow luers from the purge tubing.
11. Connect the yellow luer on the end of the purge tubing to the yellow luer on the clear
sidearm of the Impella RP
Figure 6 Connecting the Luer to the Impella RP
12. When the controller detects that the luer is connected, it automatically begins priming the
purge lumen.
SECURE THE PURGE TUBING
1. To complete the setup, connect the purge tubing to the white connector cable by pushing
the purge tubing into the clips attached to the white connector cable as shown in Figure 7.
Figure 7 Connecting the Purge Tubing to the Connector Cable
®
Catheter as shown in Figure 6.
®
Catheter
User Manual