Severity
Alarm Header
Air in Purge System
Battery Critically Low
Battery Failure
Battery Failure
Battery Temperature
High
Complete Procedure
Controller Failure
Controller Failure
Emergency Shutdown
Imminent
Impella Disconnected
Impella Failure
Impella Stopped
Impella Stopped
22
ALARM MESSAGE SUMMARY
Table 4 briefly describes all of the alarm messages that may appear on the Automated Impella
Controller when used with the Impella RP
Table 4 Automated Impella
Action
The purge system has stopped. Initiate
De-air Tool and follow the instructions to
remove the air from the system.
Plug controller into AC power.
Plug controller into AC power.
1. Plug controller into AC power.
2. Press switch located on the underside
of the controller.
3. Switch to backup controller.
Switch to backup controller.
1. Follow the steps on the screen or
2. Exit the procedure
Switch to backup controller.
The purge system has stopped, switch to
back-up controller.
Release ON/OFF push button.
1. Check cable connection to console.
2. Check Impella connection to cable.
Replace Impella.
1. Replace white connector cable.
2. Switch to backup controller.
3. Replace Impella pump.
1. Restart Impella.
2. Replace Impella after 3rd unsuccessful
restart attempt.
Catheter.
®
®
Controller Alarm Messages
Cause
There is air in the purge tubing.
Battery power has 15% remaining
capacity.
One of the batteries has failed.
A battery switch is turned off or
there is a malfunction of the switch.
Battery temperature is greater than
60°C.
Complete Procedure serious alarm
(yellow; see next page) is active and
the user has not responded for an
additional 2 minutes.
There is a problem with the controller
electronics.
The controller has detected a purge
pressure sensor defect and has
stopped the purge system.
Power switch pressed for 15 seconds
while Impella RP
Catheter still
®
connected.
Running Impella RP
Catheter
®
disconnected.
There is a problem with the Impella
RP
Catheter motor.
®
There is a problem with the
electronics.
There may be a mechanical or
electrical problem in the Impella RP
Catheter.