- press [F+P4]
- to exit the function, press [F+P4]
CALIBRATION PROGRAMS
Table T7 lists the cases in which machine calibration should be carried out. Calibration
must be carried out whenever one or more of the conditions listed occur.
Table T7: Conditions for carrying out machine calibration
Condition
When the machine is installed at the final user's
site
When the electronic circuit board is replaced
When a mechanical part linked to the pick-up
signal (pick-up, pick-up compression springs,
suspension unit + shaft) is replaced
When the pick-up compression spring
adjustment is modified
When the encoder disc is replaced
When the machine does not provide coherent
wheel balancing results
When there are large, constant variations in
the environmental humidity and temperature
conditions (for example seasonal changes)
The machine requires two independent calibrations:
• Calibration for the CAR/SUV Wheel Type (calibration is the same for both types of
wheel);
• Calibration for TRUCK Wheel Type.
To operate properly, the machine must be calibrated. Calibration allows storing the
mechanical and electrical parameters specific to each machine so as to provide the best
balancing results.
User manual ProLine Truck 150
+
keys to interrupt the procedure;
+
keys.
Who must perform
Status
Mandatory
Technical Service
Mandatory
Technical Service
Mandatory
Technical Service
Mandatory
Technical Service
Mandatory
Technical Service
Final user and/or
Recommended
Technical Service
Final user and/or
Recommended
Technical Service
it
UK
113