Table 3: Guidance & Declaration - electromagnetic immunity concerning Con-
ducted RF & Radiated RF
The model CanalPro X-Move is intended for use in the electromagnetic environ-
ment specified below. The customer or the user of the model CanalPro X-Move
should assure that it is used in such an environment.
Immunity test
Conducted RF
lEC 61000-4-6
Conducted RF
lEC 61000-4-6
Radiated RF
lEC 61000-4-3
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic prop-
agation is affected by absorption and reflection from structures, objects and
people.
34
IEC 60601
Compliance
test level
3 Vrms
150 kHz to
80 MHz
6 Vrms
3V
ISM frequen-
6V
cy band
3V/m
3 V/m
80 MHz to
2.7 GHz
Electromagnetic environment -
level
Portable and mobile RF communi-
cations equipment should be used
no closer to any part of the models
CanalPro X-Move, including cables,
than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation
distance
d=1.2×P
1/2
d=2×P
d=1.2×P
d=2.3×P
where P is the maximum output pow-
er rating of the transmitter in watts (W)
according to the transmitter manu-
facturer and d Is the recommended
separation
distance in meters (m).
Field strengths from fixed RF transmit-
ters, as determined by an electromag-
netic site survey (a.)
should be less than the compliance
level in each frequency range (b.)
Interference may occur in the vicinity
of equipment marked with the follow-
ing symbol:
guidance
1/2
1/2
80 MHz to 800 MHz
1/2
800 MHz to 2.7 GHz